| Home > Astronomy > Detectors > MITLL3 | 
| 
 
 
Operational Characteristics
 | 
| Speed | Slow | Fast | 
| Bias(ADU) | 6451 | 3450 | 
| Gain(e/ADU) | 0.67 | 1.37 | 
| Noise(e) | 2.3 | 2.8 | 
| Noise (ADU) | 3.4 | 2.0 | 
| Linear to(ADU) {also see graphs below} | 65K | 60K | 
| Readout time(s) | 60 | 39 | 
 
         
        Windowing is available.
Binning has been tested up to 4 x 4.
| Vertical CTE | 0.999997 | 
| Horizontal CTE | 0.999994 | 
| X Pixel size | 15 microns | 
| Y Pixel size | 15 microns | 
| X size in pixels of digitised area | 2148 | 
| Y size in pixels of digitised area | 4200 | 
| X size of useful imaging area | 2043 | 
| Y size of useful imaging area | 4098 | 
| X start of useful imaging area | 17 | 
| Ystart of useful imaging area | 1 | 
| LN2 capacity of cryostat | 2.5 litres | 
| Cryostat window thickness | 4 mm | 
| Distance from window to CCD | 11 mm measured optically | 
| Cryostat window size | 35 mm x 70 mm | 
| Flooding type | none | 
| Operating temperature | 163K | 
| Preferred amplifier | B (right hand side) | 
 
      
               
| Dark current | ? | 
| Cosmetics | Bright trap at [823,2592], Dark frame showing window to be used with
  Oasis There is a defect in column 712. | 
 
      | Top | Back | 
| 
 |