| Home > Astronomy > Detectors > Tek1 |
|
Warning : these figures refer to Tek1 when it was used with a Dutch controller and are for reference only. Tek1 Parameters
Device Identification
|
| Chip name | Tek1 |
| Year of manufacture | 1991 |
| Serial number | 1347BR12-02 |
| First light on La Palma | March 1992 |
| Description | Thinned AR coated |
Operational Characteristics
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Vertical CTE | 0.99999 |
| Horizontal CTE | 0.99999 |
| Pixel Scale (ISIS) | 0.33 arc/pixel |
| X Pixel size | 24 microns |
| Y Pixel size | 24 microns |
| X size in pixels of digitised area | 1124 |
| Y size in pixels of digitised area | 1124 |
| X size of useful imaging area | 1024 |
| Y size of useful imaging area | 1024 |
| X start of useful imaging area | 50 |
| Ystart of useful imaging area | 0 |
| LN2 capacity of cryostat | 2.5 litres |
| Cryostat window thickness | 4 mm |
| Distance from window to CCD | |
| Cryostat window size | 40mm x 40mm |
| Flooding type | none |
| Operating temperature | 173K |
| Preferred amplifier | B |
| Network name | CCD2 |
| Anti-blooming available | no |
| Dark current | 9 e/hour |
| Cosmetics | 11 small traps |
| Top | Back |
|