Home > Astronomy > Wide Field Camera > CCD Quality Control

CCD Quality Control

The bias value, readout noise, gain, and in the future also the linearity of the WFC CCDs are monitored by the Support Astronomers. SA, please update this table in any D nights and, if possible, also at the beginning of each observing run with a visiting astronomer. Report any anomalous value to the Operations team via a Fault report.

Date
Readout
Speed
Bias (ADU)
Readout Noise (ADU)
Gain (e-/ADU)
CCD1
CCD2
CCD3
CCD4
CCD1
CCD2
CCD3
CCD4
CCD1
CCD2
CCD3
CCD4
NOMINAL
(Jun 2013)
slow 2103
2218
2239
2176
4.5
2.7
3.5
2.9
1.2
1.7
1.2
1.6
2006-10-31

1555
1503
1552
1571
2.9
4.3
4.9
3.3




2006-11-28

767
813
718
725
2.9
2.2
3.2
2.5
2.6
3.2
2.5
3.1
2006-12-12

749
804
712
722
2.8
2.0
3.1
2.5
2.8
3.2
2.6
3.2
2006-12-24

729
793
712
724
3.0
2.2
3.1
2.4




2007-01-03

726
784
715
728
2.8
2.2
3.2
2.5




2007-01-04

723
790
711
724
2.8
2.0
3.0
2.3
2.8
3.2
2.5
3.1
2007-04-10

1944
1931
2016
1934
2.8
2.1
2.9
2.4
2.6
3.2
2.5
3.1
2007-05-10
fast
1917
1917
1976
1910
3.4
2.7
3.5
3.0
2.6
3.5
2.6
3.3
2007-05-10
slow
2218
2246
2288
2219
4.5
3.2
4.5
3.4
1.3
1.7
1.2
1.6
2007-05-17
fast
1917
1922
1976
1910
3.7
2.7
3.6
2.8
2.6
3.5
2.5
3.3
2007-05-17
slow
2218
2251
2290
2222
4.2
3.5
4.6
3.5
1.3
1.7
1.2
1.6
2007-06-07
fast
1908
1918
1987
1917








2007-07-14
fast
1896
1915
1960
1910
2.8
2.2
2.6
2.5
2.6
3.5
2.5
3.3
2007-07-16
slow
2220
2265
2292
2238
4.4
3.3
4.0
3.4
1.3
1.7
1.2
1.6
2007-11-07
fast
1897
1912
1965
1900
2.9
2.3
2.7
2.6




2007-11-07
slow
2206
2245
2281
2213
4.3
3.0
3.8
2.2
1.3
1.7
1.2
1.6
2008-01-16
fast
1894
1908
1960
1899
2.9
2.5
2.6
2.6
2.6
3.4
2.4
3.2
2008-01-16
slow
2208
2239
2280
2215
4.9
3.6
4.0
3.3
1.3
1.7
1.2
1.5
2008-01-22
fast
1893
1908
1960
1900
2.9
2.4
2.7
2.7
2.5
3.4
2.3
3.2
2008-01-22
slow
2210
2244
2281
2220
4.9
3.9
3.9
4.4
1.3
1.7
1.2
1.6
2008-11-26
slow




3.3
2.7
2.8
2.5




2011-06-17
slow
2059
2201
2204
2156
3.6
2.5
3.5
2.7
1.0
1.3
0.9
1.2
2011-06-17
fast
1838
1908
1949
1895
2.6
2.0
2.7
2.4
2.4
3.1
2.2
2.9
2011-07-03
slow
2072
2219
2216
2167
3.7
2.6
3.6
2.8
1.1
1.5
1.0
1.4
2011-07-03
fast
1853
1925
1959
1906
2.7
2.0
2.7
2.4
2.7
3.6
2.6
3.4
2011-09-10
slow
2072
2226
2217
2165
3.6
2.5
3.6
2.8
1.3
1.7
1.2
1.6
2011-09-10
fast
1851
1931
1959
1904
2.4
2.0
2.6
2.3
2.4
3.2
2.3
3.0
2011-11-11
slow
2046
2169
2221
2141
8.0
10.6
7.8
5.9
1.2
1.6
1.1
1.5
2011-11-11
fast
1821
1873
1953
1882
7.1
6.0
7.6
5.5
1.5
2.0
1.4
1.9
2012-09-29
slow
2107
2228
2242
2164
4.6
2.8
3.9
3.1
1.1
1.5
1.1
1.5
2012-09-29
fast
1906
1943
1986
1911
3.0
2.1
2.7
2.2
0.6
0.8
0.5
0.7
2012-11-05
slow
2117
2217
2288
2197
4.5
2.7
3.8
3.0
1.2
1.7
1.3
1.6
2012-11-05
fast
1912
1940
1982
1910
3.3
2.2
2.8
2.3
1.9
2.5
1.8
2.4
2013-02-01
slow
2060
2172
2217
2148
6.6
4.0
4.3
3.6
1.2
1.6
1.2
1.5
2013-02-01
fast
1881
1907
1977
1908
3.9
2.3
2.6
2.6
2.1
2.8
2.1
2.7
2013-04-05
slow
2092
2195
2231
2168
4.6
2.7
3.3
2.6




2013-04-05
fast
1897
1915
1977
1910
2.9
1.9
2.2
1.9




2013-06-03
slow
2103
2218
2239
2176
4.5
2.7
3.5
2.9
1.2
1.7
1.2
1.6
2013-06-03
fast
1906
1930
1981
1921
3.2
2.3
2.5
2.2
2.5
3.4
2.5
3.3