Autofib 2/WYFFOS Telescope Instrument Setup Procedure
Section 5 : Installing AF2 part 2/2
This sections gives instructions for the second part of the installation
procedure.
Fit Ancillary items at top end
- Fit the swans neck.
- Fit A&G controller support frame and install SDSU controller.
- Place the top end ring working platform in place and bolt down
- Install TV camera system with AG1. Make sure the camera sits against
the two reference faces. Optimum focus is at 5.42 mm on the micrometer.
Cable-UP Top end.
- Prime Focus Unit to AF2 Robot, AG2 acquisition, AG1 tv/autoguiding
- Cable up Autoguider. Fibre connects to prime 1.
- Connections to the AG1 TV and interface
Connect up Science and Guide TV fibres.
-
Clear area and ensure only change staff are present.
-
Lay out all the fibres on the floor.
-
Run the red science fibres from AF2 up to the topring
-
Run the grey guide fibre to the AG1 TV.
-
Great care must be take when handling the fibres.
-
Using cable ties secure all the fibres at the end of the swan's neck. Ensure the cable ties are
on the metal part not on the grey mesh.
- Very important. Move rotator locally to +/- limits. Check fibres twist
OK. When A&G is at the top the rotator PA should be 90 degrees.
- Remove top end ring working platform.
Balance Telescope.
-
Drive weights to the bottom.
-
Paper calculation on moments removed and added.
-
Release telescope ties when you are sure telescope is near to balance and
slightly bottom end heavy.
-
Check top to bottom
-
Check across cube
-
Check top to bottom
This concludes the Installation of AF2 on to the telescope. The final
thing to do is to check that the instrument works, see Section
6