Astronomy

Logo
Home · Search · About ING · Astronomy · Public Information · Engineering

The OASIS Obstool User's Guide


  1. Introduction
  2. Quick Reference Guide
  3. Starting up the OASIS Obstool
  4. Setting the demand values: Set Demand
  5. Configuring the mechanisms: Configure
  6. Manual adjustments to the configuration: Adjust
  7. Taking integrations: Expose
  8. Summary of available setups
    1. Spectral Configurations
    2. Spatial Samplings
    3. Observing Modes
  9. Discussion of when to override the defaults
    1. Dichroic
    2. Filter Focus
    3. Iris
  10. Examples
  11. Shutting down the OASIS Obstool
Back to the Oasis internal pages

1. Introduction



The OASIS Obstool provides a simple, efficient and accurate means to set up OASIS observations. By selecting the (i) spectral configuration, (ii) spatial sampling, and (iii) observing mode (e.g. an arc frame) desired, one can use the tool to automatically configure the 21 critical mechanisms (i) on the AO bench and (ii) within OASIS itself.

This webpage gives a step-by-step guide to using the OASIS Obstool.

Back to the top 

2. Quick Reference Guide



This section gives an overview of how to use the OASIS Obstool and is intended as a quick reference guide. Detailed instructions on each step are given in the rest of this user guide.

  1. Bring up the Instrument Control GUI.
  2. Select the OBSTOOL tab.
  3. In the Set Demand section, use the drop down menus to select the Spectral config, Spatial sampling and Observing mode you require.
  4. In the Configure section, click in the tickboxes to select the mechanisms that you wish to configure.
  5. Click on Configure Selected Items.
  6. Once the mechanisms have stopped moving, both OASIS and the AO bench will be correctly configured for the spectral configuration, spatial sampling and observing mode that you requested.
  7. Take your exposure using the appropriate ULTRADAS command.


Back to the top 

3. Starting up the OASIS Obstool



There are a number of ways to start the OASIS software and obtain the Obstool GUI. For normal night time observing, however, you should refer to Situation 1.

The GUI that you should now have before you has six tabs. The separate tabs are:
  1. INGRID - This is the observer/engineering interface for the infrared science camera INGRID.
  2. USP - This is the engineering interface for the AO bench mechanisms.
  3. OASIS - This is the engineering interface for the optical integral field spectrograph OASIS.
  4. OBSTOOL - This is the observer interface for OASIS observations.
  5. TCS - This gives the status of the TCS (telescope control system).
  6. SEEMON - This gives the seeing data from the seeing monitor Robodimm.
The six tabs are displayed below. (Aside: Note that not all the grabs were taken at the same time, hence some of the status fields do not match from tab to tab.)

INGRID tab
USP tab
OASIS tab Obstool tab
TCS tab Seemon tab

Click on images to enlarge

If you find the status of (i) the dichroic (on the USP/Obstool tabs) or (ii) the OASIS mechanisms (on the OASIS/Obstool tabs) to be in an error state:

It is likely that another session of the control software is already running, perhaps on a different machine. In this case:

1. Identify which machine the software is already running on. On each possible machine (probably taurus, lpss94 or lpss39) type:
SYS> ps -ef | grep java
and search for the various servers associated with INGRID, USP, OASIS, Obstool, TCS and SEEMON.

2. Once you have found the servers running (e.g. see here) you can cleanly shut them down by typing:
SYS> shutdownoasisserver
on the relevant machine.

3. The items that were in an error state on your GUI should recover their status.

If you find the TCS information missing (on the TCS tab):

1. One reason could be that the observing system hasn't been started (properly) on the machine that you are using. The SYS> startobssys command initiates a couple of communication tasks that are vital for the TCS information to be accessible. If you need this information you will need to (re)start the observing system.

2. It is possible that one of the TCS communication tasks has failed. If there is (or has been) a problem with the OCSCorbaDramaInterface, then:

  1. Type SYS> ps -Aef | grep OCSCor
  2. Kill the process if it exists.
  3. Restart the process using SYS> OCSCorbaDramaInterface &

Back to the top 

4. Setting the demand values: Set Demand


 
You are now ready to use the Obstool to prepare an OASIS observation. The first step is to define the OASIS configuration you require and the observing mode that you will be using:
Obstool interface
Back to the top 

5. Configuring the mechanisms: Configure



Now that you have selected the demand values, you can configure the mechanisms:
Errors 1
Errors 2
Errors 3

Click on images to enlarge

Back to the top 

6. Manual adjustments to the configuration: Adjust



In the previous section, we discussed the five groups of mechanisms that are configurable with the Obstool. All the OASIS mechanisms formed one group and most of the AO bench mechanisms were included in another. However three mechanisms remained in their own individual groups (the dichroic, the filter focus and the iris), with their status and control treated separately.

The three demand values set earlier uniquely define the position of all the OASIS mechanisms and the majority of the AO bench mechanisms. On the other hand, although one can assign default values to the positioning of the dichroic, the filter focus and the iris, there are a number of reasons why you might wish to overrule these default settings. To do this, you can use the Adjust section of the Obstool GUI.

Adjust dichroic
Adjust iris
Adjust FFC

Click on images to enlarge

Back to the top

7. Taking exposures: Expose



Now that you have correctly configured both OASIS and all the AO mechanisms on the bench, you are ready to take integrations. In the near future, you will be able to launch these directly from the Obstool GUI. In the meanwhile it is still necessary to use the command line interface of the observing system. E.g.:

SYS> run oasis 300 "BD+284211"
SYS> multrun oasis 3 300 "BD+284211"
SYS> bias oasis
SYS> multbias oasis 3
SYS> dark oasis 300
SYS> multdark oasis 3 300

Note that the special bias and dark commands are required for bias/dark frames as they prevent the camera shutter from being opened. However it is not necessary to replace the run command with flat, sky, arc, and pupil while taking such exposures. The exposure type is automatically recorded in the image header using an OAFCLASS header, and is deduced from the observing mode that you requested in the Set Demand section. The actual values are represented as integers (which are recognised by the data reduction package XOasis), and the corresponding mappings can be found here.

The Obstool also generates an OACONFIG header, which contains the current configuration of OASIS (e.g. MR661_26). Note that this value corresponds to the Oasis status reported in the first field of the Configure section.

Back to the top

8. Summary of available setups



In this section we summarise the OASIS configurations that are currently implemented in the Obstool.

Spectral Configurations

OASIS offers fifteen spectral configurations, covering a variety of spectral ranges and resolutions. Their names have a five letter syntax; the first two letters indicate the resolution (LR, MR and HR for low, medium and high resolution) while the last three numbers give the central wavelength in nm. For example, the MR661 configuration is a medium resolution configuration centred at 661nm.

The filter wheel in OASIS is only capable of holding 12 filters, whereas each of the 15 spectral configuration has a unique filter. This means that at any one time, three filters (and therefore three configurations) will not be available. One can determine the 12 configurations that are currently offered from the drop down menu.

Spectral configs

Click on image to enlarge

Spatial Samplings

OASIS offers four spatial samplings (and fields of view) through the use of four different enlargers. The spatial sampling is usually referred to in terms of the sampling per lenslet, and the available values are 0.09"/lens, 0.14"/lens, 0.26"/lens and 0.42"/lens. Note that the naming convention for the full OASIS configuration (i.e. both spectral and spatial configuration) also uses these values, e.g. MR661_26 corresponds to the MR661 spectral configuration combined with 0.26"/lens spatial sampling.

All four spatial samplings can currently be configured from the Obstool using the drop down menu. (The 0.42"/pix sampling is not currently used for science as it has not been commissioned due to severe vignetting.)

Spatial sampling

Click on image to enlarge

Observing Modes

16 OASIS observing modes are currently offered. Here we give a brief summary:

Observing Mode
When?
AO Bench Setup
OASIS Mode
Why?
Spectroscopy
On-sky science
Light from telescope to OASIS
Spectroscopy
On-sky science
Imaging
On-sky acquisition
(with OASIS)
Light from telescope to OASIS
Imaging
On-sky imaging
Mainly for finetuning the acquisition
Acq cam (for OASIS spectro)
On-sky acquisition
Light from telescope to AG3
Spectroscopy AG3 acquisition
(OASIS prepared for spectroscopy)
Acq cam (for OASIS imaging)
On-sky acquisition
Light from telescope to AG3
Imaging AG3 acquisition
(OASIS prepared for imaging)
Arc (Neon)
On/Off-sky calibration
Light from neon lamp to OASIS Spectroscopy Arc frames for mask creation / wavelength cal.
(AG3 acquisition also possible)
Arc (Helium)
On/Off-sky calibration Light from helium lamp to OASIS Spectroscopy Arc frames for mask creation / wavelength cal.
(AG3 acquisition also possible)
Bias
Off-sky calibration
N/A (but switches lamps off)
N/A
Bias frames for CCD reduction
Continuum
Off-sky calibration Light from continuum lamp to OASIS
Spectroscopy Continuum frames for mask creation
Dark
Off-sky calibration N/A
N/A
Dark frames for CCD reduction
Dome
Off-sky calibration Light from telescope to OASIS
Spectroscopy Dome frames for flat fielding
Mu-pupils
Off-sky calibration Light from continuum lamp to OASIS Micropupils
Micropupil frames for mask creation
Sky
On-sky calibration
Light from sky to OASIS
Spectroscopy Afternoon (twilight) flats for flat fielding
White light flat
Afternoon Setup
Light from simplex lamp to AG7
N/A
Performing a white light flat on the WFS
Simplex (AG)
Afternoon Setup
Light from simplex lamp to AG7
N/A
Checking a pinhole image on AG7
Simplex (OASIS)
Afternoon Setup
Light from simplex lamp to OASIS
Imaging Checking a pinhole image on OASIS
Shutdown
End of night
Lamps off, bench fixed, (dichroic parked)
Shut
Safely shutting down the system

AO Bench Setup: Note that the Obstool ensures that (i) the table is fixed if necessary, (ii) the NCU is in the appropriate position, (iii) the correct lamp (if any) is on and set to an appropriate intensity, (iv) unused lamps are switched off, (v) the correct dichroic is selected, (vi) the Oasis and Simplex flats are correctly deployed, and (vii) the filter focus corrector is set to a suitable value, all for the Oasis configuration and observing mode requested.

Oasis setup: Oasis is configured according to (i) the spectral and spatial configuration requested and (ii) the observing mode. Column 4 summarises whether the mechanisms are configured for OASIS's spectroscopic, imaging or micropupil mode (in the desired spectral/spatial configuration). Note that there is also a special "Shut" mode, for which the Entrance Wheel shutter must be deployed (but the position of the remaining OASIS mechanisms is irrelevant).

Observing mode

Click on image to enlarge

Back to the top

9. Discussion of when to override the defaults



Earlier we discussed how it was possible to use the Adjust section of the Obstool to change the positions of the dichroic, filter focus, and irises from their default values. Here we outline possible reasons why this might be done:

Dichroic

For each OASIS configuration there is a default dichroic assigned to it. This dichroic must have a large enough bandwidth in order to enclose the bandwidth of the associated OASIS filter, yet should be as narrow as possible to ensure that as much light as possible goes to the WFS (wavefront sensor) of the AO system. Possible reasons to change from the default position are:

(i) The MR807 configuration can be used with both dichroic #3 and #4. While the current default is to use dichroic #3, in some cases the observations might be better suited to using #4.

(ii) Nine of the configurations that by default use dichroic #1 or #2, could also be employed with dichroic #5. This could be beneficial if one wishes to use an extremely bright AO guide star, as it would reduce the magnitude of the star on the wavefront sensor and help prevent the latter from becoming saturated.

(iii) If you are observing a bright object (and thus are not photon starved) but require a very good AO correction, it may be beneficial to use the 50:50 dichroic instead of the default OASIS dichroic. The 50:50 dichroic sends half of the light (at all wavelengths) to OASIS, and the remainder to the WFS. Depending on the configuration and dichroic that you are using, you may get an increase in flux on the WFS by adopting the 50:50 dichroic, and can therefore improve the performance of the AO correction. Note however that the flux reaching OASIS will be reduced by a factor of ~2.

(iv) During the daytime setup for an OASIS run, the support astronomer (SA) must flatten the DM (deformable mirror). A couple of observing modes [Simplex (AG) and Simplex (OASIS)] have been added to allow the SA to both perform the flattening, and assess the results, using an illuminated pinhole. By default the Simplex (AG) mode adopts the 50:50 dichroic, while Simplex (OASIS) uses the default dichroic for the OASIS configuration requested. However it is possible that the SA may wish to "simplex" using a different dichroic; similarly they may want to check the image quality on OASIS using Simplex (OASIS) with the 50:50 dichroic.

Filter Focus

The filter focus corrector mechanism is used to apply focus offsets when one switches between (i) different OASIS filters and (ii) between OASIS and AG7 (the simplex camera). While default offsets have been included in the Obstool, the observer may wish to check that these offsets provide the sharpest focus and, if necessary, adopt a different value.

Iris

Many of the observing modes use one of the four calibration lamps (Ne, He, continuum or simplex). "Spectral irises" are used to provide the large dynamic range in intensity that is required for the many OASIS configurations. Where relevant, a default iris setting has been established for every OASIS configuration and observing mode to ensure that an exposure of a few seconds will give ~30-40,000 counts. (Aside: Note that longer exposure times are needed with the Neon arc lamp for the bluer configurations; in future a new Argon lamp will be used instead.)

You may need to adjust the default iris settings for the following reasons:

(i) The actual lamps are brighter or fainter than when the default values were measured (in particular when the lamp is about to fail).
(ii) The lamp has just been replaced. Note that if the lamps appear a lot fainter than they previously were, it could indicate a problem with their alignment; you should ask Olivier Martin or Tom Gregory to check this.
(iii) The dichroic you are using is not the default one (see here).
(iv) There is some hysteresis in the iris mechanisms, which may have an effect particularly for small irises.
(v) In the short term, the iris settings have only been measured for the 0.26" spatial sampling. It may be possible to simply adjust the exposure times to compensate for the increase/reduction in flux when using coarser/finer samplings, but in some cases it may be necessary to adjust the iris.

Please report any major changes in the required iris settings to Chris Benn or Sam Rix.

Note that the default iris settings may be found to be unsuitable simply because the iris mechanism needs initialising. Go to the USP tab of the Oasis software, find the appropriate iris ("Spectral" for the Ne or He lamp, "Continuum" for the continuum lamp, and "Simplex" for the simplex lamp), and select "Initialise" from the drop down menu. The mechanism will take a few minutes to initialise, then you can move it again to the default setting.

Back to the top

10. Examples



In this section we illustrate the use of the OASIS Obstool by using an example sequence of events that demonstrate the key features. (Click on the images to enlarge.)

1. The image below shows the Obstool after it has been successfully configured for an MR516_14 continuum exposure. All the fields in the Configure section are in green, meaning that their status satisfies the 3 demand values in the Set Demand section.

Example 1

2. We now decide to take a arc exposure, using the Neon lamp, with the MR661_26 configuration. First we change the 3 demand values. We can now compare the status of the five fields in the Configure section to their previous states. First note that the status of OASIS, the dichroic and the filter focus remains unchanged. However, as our demand observing mode has changed from continuum to arc, the iris status displayed has changed from that of the continuum iris to that of the spectral iris. Finally the "Naomi other" status has changed from "CONFIGURED OK" to "NOT CONFIGURED". More strikingly, the colour coding of all the fields has changed from green to pink, indicating that none of the groups of mechanisms are correctly configured for the new demand values.

Example 2

3. Next we ensure that all five tickboxes are ticked, and click on Configure Selected Items. The appropriate mechanisms are moved to the correct position and their status is shown in blue while they are moving.

Moving

Once the mechanisms have stopped moving, and they are correctly positioned for the three demand values, their status is shown in green.

Example 3

Note that if you wish to see the corresponding status of the individual OASIS mechanisms you can click on the OASIS tab. Normally this is not necessary, but it can be useful if one of the mechanisms has gone into an error state.

Example 3a

Similarly the status of the individual bench mechanisms can be checked by clicking on the USP tab.

Example 3b

4. Having taken a test exposure, we might decide that the arc lamp is a little too bright. We can reduce its intensity by stopping down the spectral iris. We go to the Obstool's Adjust field, select Edit value, type in -310, and hit enter. The spectral iris mechanism moves to the new demand position, and the final status is displayed in pink to remind us that this is not the default position for the MR661_26 Neon arc configuration. We can now take our arc exposure.

Example 4

5. We now decide to take another Neon arc, but with the MR516_26 configuration. Selecting the new demand values, we find that the "Naomi other" mechanisms are correctly configured, but the positions of the OASIS, dichroic, filter focus and iris mechanisms are incorrect.

Example 5

6. We decide to configure all the mechanisms except the filter focus, as the latter is not so important when taking arc frames. To deselect the filter focus we click once on its tickbox. Note that we do not need to bother unticking the "Naomi other" box even though it is correctly configured.

Example 6

7. We click on Configure Selected Items and the relevant mechanisms are repositioned. The filter focus remains in its current position and is colour coded pink to remind us that it is not in the default position for an MR516_26 Neon arc frame. Note that the spectral iris has been reconfigured i.e. if you put a tickmark by the iris field, and click on Configure Selected Items, the mechanism will be moved to its default position regardless of the value previously selected in the Adjust section.

Example 7

Back to the top

11. Shutting down the OASIS Obstool


At the end of the night (or the end of daytime work), you should configure both OASIS and the AO bench into the "Shutdown" mode:

You do not need to close down the OASIS control software if you are using the main observing system (taurus) and OASIS was configured into the system. However if you are performing daytime work (particularly on a different machine) you should always shut down the software once you have finished with it. If you do not do this, problems will arise when someone else tries to run the software (possibly from a different machine). To cleanly shut down the software, type:

Back to the top


ING Logo

Last Updated: 18 August 2006
Samantha Rix, srix@ing.iac.es