Starting the computer system
-
Log on to the main system terminal lpvs1 as observer. Check the password
at the whiteboard.
-
We will see a banner asking which system computer to use, LPVF or LPVG,
check the whiteboard for the right.
-
When asked if you want to start the observing system reply YES.
-
At the VMS command prompt, type ICL.
-
Answer the observer name and the PATT reference questions.
-
When ask if you want to use the current setup, select the option with WYFFOS
using INTEGRAL.
-
Answer YES to the question about loading the system.
Changing gratings
-
1.
-
To change gratings on WYFFOS you must use wchange at the ICS.
-
2.
-
Go up to the GHRIL
-
3.
-
open access panels to the grating mount
-
4.
-
press the clamps control botton on the grating mount, which is situated
facing the user. Press until red is alight.
-
5.
-
Take out the actual grating in WYFFOS
-
6.
-
mount the new grating with the label on its back right way up.
-
7.
-
Press again the clamp control botton to fix the grating clamps.
-
8.
-
Do not forget to update the new grating in the ICS with wchange
command.
At the change position (-25.0), a central wavelength, of 7810 Å
will be obtained. Any other can be selected with the WYFFOS command wcenwave.
Changing gratings should normally be done by your support astronomer or
ING technical staff.
WYFFOS grating data
SA must to check the grating angle offset. This parameter enable
correct calculation of central wavelength. If not properly set, the read
central wavelength will not be correct. To check and change it type:
ICL> wchange
option? [0] 3
:
:
Current grating angle offset is xxxx mdeg (If different from
6170 mdeg, change it!)
New value? [0] > 6170
In the table 1 is shown for each grating
leads the central wavelength 6328 Å(He-Ne Laser).
Table1: Grating angle for obtaining
a central wavelength of 6328 Å
| Grating |
Angle (o)
|
| R300B |
- 2.44
|
| R316R |
-2.73
|
| R600R |
-7.95
|
| R1200R |
-19.35
|
Slit traslation (height) adjustment
The slit sould be enough centred on the CCD in order to include all the
fibers (of each bundle) on the CCD. This step could be ckeck only with
the SB1 bundle.
-
1.
-
Setup spectrograph to obtain a flat:
-
Turn on the white lamp: ILA W
-
Select STD1bundle: ISP STD1 CALIBRATION
-
Select a central wavelength: WCENWAVE 5000
-
Take a short exposure: RUN INTEGRAL 2 ``test STD1''
2.
-
Use int_slit to check if all fibers are on the CCD:
-
type integral to access to the Integral package within IRAF
-
epar int_slit and input parameters
-
Ckeck that 205(for STD1) fibers are present at different lines on the CCD
3.
-
If 205 fibers are not present at any line, traslate in height the
slit:
-
a.
-
ICL> wslitt xxxx (microns, from -4500 to 4500) ->This command sets
the WYFFOS fibre slit to a specified position in microns using the slit
traslation drive. To get into the CCD one fiber (from the left) traslate
the slit by -1000 microns.
-
Focusing of the spectrograph
The CCD surface in WYFFOS is not flat - it is cut as a section from a spherical
surface, so it is not possible to position the CCD to give (near) zero
Hartmann shifts over the complete imaging area. The compromise was to adjust
the micrometers to give a positive Hartmann shift at the centre of the
CCD, and a negative (balanced) shift around the edge.
The previous steps to be carried on are:
-
mount the R1200B grating and select a central wavelength of ~5500
Å
ICL> wcenwave 5500
-
mount the BG39 filter and put in the beam
ICL> wfiltA in
-
put the STD3 in the calibration position
ICL> ISP STD3 CALIBRATION
-
Turn on the spectral calibration lamps (CuAr+CuNe)
ICL> ILA CuAr on// ICL> ILA CuNe on
-
Set WYFFOS to the start focus position (nominal=0)
ICL> wslitf 0
This nominal position is with the filter BG38 in the beam. Removing
the filter applies an offset of about -1000 units
-
close the right Hartmann shutter
ICL> whart R
-
Take an exposure
-
close the left Hartmann shutter
ICL> whart L
-
Take other exposure
-
Select at least 9 position on the detector including an spectral line,
and write the x, y position on an ascii file
-
run the task int_focus to obtain the Hartmann shifts for each selected
position. The best focus should have values for shifts similar to the ones
shown in Figure 1. There is no serious
degradation in performance if the shifts are within 0.3 of this values,
as the resolution of the spectrograph is limited by the fibre size. At
the central position, the shift should be keep close to 0.5. This should
be normally accommodated using the slit focus: a change of 1000 microns
in slit focus corresponds to -0.4 in Hartmann shift.
Figure 1: Hartmann shifts
 |
Do not forget to open the Hartmann shutters after Hartmann test.
whart OUT
The focus can be also obtained looking for the minimum FWHM of the fibre
images at the CCD. For this:
-
1.
-
Take one arc exposure with both Hartmann shutters open.
-
2.
-
Select at least 9 position on the detector including an spectral line,
and write the x, y position on an ascii file
-
3.
-
run the task int_fwhm to obtain the Hartmann shifts for each selected
position. The best focus should have values for shifts similar to the ones
shown in Figure 2.
Figure 2: FWHM countours at the best focus
 |
Here you can find typical focus values.
Focusing and Orientation of the Acquisition Bundle
The end of the acquisition bundle in front of the TV camera should be focussed,
using the manual mechanism (see Figure 3). In principle there are two procedures,
both requiring the interaction of two people: one at the GHRIL (moving
the focus mechanism), and the other in the control room watching the TV
screen.
i) With the bundle connected to the to the SP, the inside of AG&C
box is illuminated (e.g., with a torch, a faint dome light, etc).
By inspecting the edges of the image bundle on the TV screen, the best
focus can be determined. If the focus remains unsatisfactory the following
procedure must be used (as a last resort).
ii) The bundle is dismounted from the SP (see INT-IAC-PRC001), and connected
to a dedicated focusing system. This system allows a test image (e.g.,
a number) to be seen through the bundle.
Currently, a painted mark is used to fix the bundle orientation. However,
we anticipate that before the instrument reaches common user status the
bundles will be doweled in the correct positions.
In order to aid night-time acquisition the outline of the bundle should
be marked on the TV screen.
Focusing and Orientation of the Guiding
Bundle
The end of the guiding bundle in front of the TV camera can be focussed
using the manual mechanism (see Figure 5). The same procedures as those
described for focusing the acquisition bundle should be adhered to here.
If procedure i) is used, the sharpest image of the space between the two
bundles indicates the best focus position.
As is the case with the acquisition bundle the correct orientation is
currently indicated by a painted mark, but again, we anticipate that the
bundle will be shortly doweled into position.
In the standard configuration the inner (outer) field is to the right
(left) of the TV screen.
Rotator Limits
The GHRIL rotator has hardware switches limiting its motion to slightly
less than one revolution. The TCS software limits are set about 1 degree
inside the hardware limits, constraining the rotator mount position angle
to a range of 3 to 358 degrees. The safety of integral should not
be entrusted to these mechanisms -- the rotation of the instrument should
be checked visually several times during each night, after which an assessments
of the reliability of the limits can be made.
Rotator Zero Set
Owing to the fact that the absolute encoder `wraps' at a certain
point, it is very important to zeroset the incremental encoder against
the absolute at a known physical position. That position is mount PA 180,
and it is the `park' position for integral. When the rotator is
at this position the `180' label marked on the rotator in indelible ink
is at the 12 `O' Clock position when viewed from the GHRIL bench. Before
observing with integral it is important to ensure that the rotator
is in this position (to within twenty degrees or so is adequate).
The local controls can be used to move it if necessary. The incremental
encoder can then be zeroset against the absolute in the usual way. It must
not be zeroset without a visual check that the rotator is in the required
position.
The Guide Star Server -- GSS
The GSS with configurations for WHT INTEGRALINNER and WHT INTEGRALOUTER
is selected when you log on to the GSS account and select WHT.
Control in Engineering Mode
To start the wyffos mimic, log in to lpss3 (from a xterm), as
user wyffos (username: wyffos; password: wht_wyffos)
and
type
start_eng
To start the integral mimic log in to orion.roque (from a xterm)
as user integral (username: integral; password:
wht_integral)
and type
start_eng
The low level Software Users Manual has the reference INT-IAC-MNL-001
Last updated March 2000
Begoña García-Lorenzo (Instrument specialist)
bgarcia@ing.iac.es
Ana M. Pérez-García (Deputy)
aperez@ing.iac.es