wheelcon Command


This page is part of the ING manual WHT-INGRID-2: INGRID IMPB Software Operations Manual



 
Usage: wheelcon

Category: Mechanism Command
 
Description: This command displays a list of the various optical components (filters and pupil stops) loaded into INGRID's mechanism wheels.

Note: changing INGRID's filters and pupils stops is non-trivial and not something that is expected to occur very often. In use filters and pupil stops are maintained at cryogenic temperatures. Changing them results in the need to warm up the instrument and break the instrument vacuum seal.
 

Arguments: None
 
Examples: wheelcon - display wheel status information.