Previous: Target acquisition with a multiaperture mask
Up: Observing procedure
Next: Calibration exposures
Previous Page: Further refinements to acquisition
Next Page: Calibration exposures
You need to consult this section only if you are using one of the supplied longslit masks (Section 1.2). If you are using a multiaperture mask that contains one (long) slit which also has fiducial holes custom-made for the target field, then you should use the multiaperture acquisition procedure in Section 5.3.
The procedure here is as follows:
,
) in the LEXT
coordinate frame. For example, if
-positions measured on the Tektronix
CCD (in pixels) are 567.4 for the slit and 578.2 for the target, the offset
is
arcsec.
in degrees which is positive if you
need to move the line joining the targets clockwise to align them
with the slit direction. For example, if you want to place the slit on two
targets with positions (571.9,606.9) and (570.7,315.6) ,
.
This assumes that the slit is precisely aligned with the columns of the
detector. If it is not for any reason, you will need to include this
misalignment in the angle calculation.
Note that this also assumes that the targets are
roughly symmetrically disposed about the centre of the field. If this is
not the case, a small lateral displacement might be introduced when the
angle is changed. To cure this you would need to solve for
and
together rather than independently.
as described in step (7) in the standard multiaperture acquisition sequence (Section 5.3). Note that if you BLIND_OFFSET from a pointing recalibration star (step (2) of the standard multiaperture acquisition sequence) the TWEAK is likely to be smallTWEAK
![]()
![]()
![]()
Complications may arise if the size of the required telescope offset is large because: