Previous:  Introduction.
 Up:  Requirements for new CCDs on ING Telescopes.
 Next:  INT Requirements.
 Previous Page:  Introduction.
 Next Page:  INT Requirements.
 
The Utrecht-Echelle Spectrograph,  UES and the Intermediate Dispersion
Grating Spectrograph,  ISIS (red & blue) arms can be served 
by a sensor of size ; with  
 pixels. A larger sensor could
capture  some of the UES vignetted field,  but
would not  help much  on  ISIS. A  smaller  one (
) would in fact be
sufficient for ISIS but the desired pixel size remains the same.
The Wide Field Fibre Spectrograph  WYFFOS, currently undergoing tests at
RGO,  has been designed to  accommodate a fixed-size embedded CCD. Originally  planned
for a  thin EEV 05-30 (),  we now plan to incorporate a TK1024  device
(
) as the former EEV devices are not yet available (Fri Sep  9 14:56:09 BST 1994
).
  pixels are ideal for  the standard  Autofib-Prime Focus feed
(Autofib is the automated fibre positioner being built by the University of
Durham); smaller (
) pixels would be required for a (future) f/11 feed
for this instrument.
The Prime Focus,  PF, instrument has been designed to feed a 60 mm
field-of-view ( FOV), although  the telescope can provide  140 mm. The  
PF therefore requires  a   detector, with 
 pixels.
The Low-Dispersion Survey Spectrograph, LDSS, and the imaging Fabry-Perot Spectrograph, TAURUS both have a FOV of about 30mm. These instruments ought to be served by the existing Tek heads, or any new PF device. It is not considered essential to offer a purpose-built CCD camera head for these instruments.
The Faint Object Spectrograph,  FOS2, was originally designed for a TK512 CCD,
but has an un-thinned GEC device installed. This instrument requires a  sensor
with  pixels and an area of  about 
.