Previous: GLA(NCE)/GFILE - A trial exposure
Up: CCD CONTROL COMMANDS
Next: DARK - Perform a Dark-Count exposure
Previous Page: GLA(NCE)/GFILE - A trial exposure
Next Page: DARK - Perform a Dark-Count exposure
Perform a run, always into a new file:-
Parameters
As for GLANCE, the exposure time in seconds, prompted for if omitted.
The object name can be given as a second parameter. It should be in double quotes and will be prompted for if omitted.
Note that for runs of less than 10 secs the run will not
start unless you respond to the prompt for the name.
Examples
Adam:> RUN 1000 ``LMC Abnormal Refraction''
Perform a 1000 second run looking at the LMC.
Notes
You can change your mind about the exposure time while the run is in progress by using the command TIME, or STOP at the appropriate moment.
During all but the briefest runs, the ADAM terminal is available
for displaying the RUNSTATus, PAUSEing, CONTINUEing, STOPping,
ABORTing and running application programs, but you may NOT use the TELE command during a run.
All runs are timed using the Time Service, and although the actual exposure time may not be exactly that which was requested, the actual exposure time will be known to about a millisecond accuracy and accurately recorded for you.
If the exposure time is zero, then as for GLANCE, the shutter
is not opened at all. This is used when taking
bias frames:
ADAM:> RUN 0 ``Zero Bias Frame''