Previous: Flat Fields
Up: WHAT EXTRA OBSERVATIONS ARE REQUIRED?
Previous Page: Flat Fields
Next Page: PREPARATION FOR OBSERVING

Dark counts

Strictly speaking, dark count frames should be subtracted from each observation in the reduction process (section 1.1.3). In practice, the dark-count is so low (Table 1.1) that this is rarely necessary. If it is (e.g. long exposures, very narrow-band filters, no sky counts), then dark frames must be obtained. To produce these, run procedure DARK:

Adam:> DARK 10000 (sec)

At the end of `exposure', the chip will be read out with data dumped to the next available run file.

To stop the run prematurely, press the break key, then

SEND CON: (break-in)

Adam:> STOP (stop the exposure)

Adam:> READ (readout)

Adam:> DISP (display)

Adam:> KEEP (store on disk or disk + tape)

The dark count frame for the RCA chip in particular does show some structure (Fig. 1.9) with a warm corner apparent at the readout point. Note that the DARK procedure is also a very effective way of measuring the cosmic-ray rate.

Fig. 1.9

dxc@mail.ast.cam.ac.uk