The cryostat should be filled to at least 1.2 litre capacity
initially,
and if kept at Zenith position say, gives about 16 hours hold time. If
the turntable is rotated to an angle other than
, then nitrogen
loss will occur and the hold time can be reduced to 10 or 12 hours.
In order to minimise nitrogen boil-off and give some
temperature control
the detector assembly (and temperature sensor) have an intrinsic time constant.
This explains the 2-3 hour cool-down period. A further effect is that if
the liquid nitrogen runs out completely then it may take an hour or so
before any temperature rise is detected; the temperature rise may then
be rapid.