Previous:  RU(N) - the primary exposure command
 Up:  CCD Control commands
 Next:  RESTART
 Previous Page:  RU(N) - the primary exposure command
 Next Page:  RESTART
 
Pause / continue / stop / abort an active run. For PAUSE, the timer is stopped and the shutter is closed. For CONTINUE the timer is restarted and the shutter is opened. STOP behaves exactly as though the timer period had expired. ABORT stops dead without reading out.
ParametersNone
Examples
Adam:> PAUSE Adam:> CONTINUE
Notes
The PAUSE command will be rejected if a run is not currently active.
The CONTINUE command will be rejected if a run is not currently PAUSEd.