This document is to be used during the CO2 / Snowcleaning of the WHT telescope mirrors.

CO2 cleaning of the telescope mirrors


When:

  • Currently we use snow cleaning on WHT M1 every 6-8 weeks, depending on weather and R% measurements.
  • If the Internal Humidity reads more than 40% do not Snowclean that day, to avoid condensation stains on the mirror.
  • Casual observation also suggests that when RH% is below 20, snowflakes do not stick as much to the mirror surface, which may reduce the effectiveness of the cleaning.
  • Power down the mirror temperature sensor amplifier before snowcleaning. The switch is located behind the Sauron PC in the control room.
  • The mirror now has an earth cable connecting the edge of the reflective surface to ground, which prevents damage to the temperature sensor by static.

    Material:

    For the Snowcleaning you need:

    Safety:

  • Before parking at AP3, the fixed end of the Fall Arrestor must be secured to the hook above the broken CASS port and the extesible lead trained out between the spider vanes. Finally place the clamp at the end of the extensible lead over the ladder hooks on the opposite side of the cube. When opening the mirror petals, one of these will slightly brush the spool housing of the fall arrestor, but should not trap it when fully open.

  • Important to know:

    Snow cleaning is done to remove the dust on the mirrors. The slow flow of snowflakes pickup the dust and falls down from the mirror.

    If you use a high pressure you might damage the mirror when hard snowflakes bounce against the soft aluminium and cause pits in the mirror coating.

    The uSMS reflectometer head is sensitive to temperature. It should be left in the dome for some time, preferably overnight, to adjust to ambient temperature. The unit can be put on the electrical charger during this time.

    Procedure:

    1. Secure the Fall Arrestor as decribed above.
    2. Make reflectivity measurements at each of the ports on the mirror petals.
    3. Park the telescope at AP3 (WHT) or Access Park (INT) and open the mirror covers
    4. Alternatively, Reflectivity can be measured at park position, but only in the bottom half, in the case of the WHT mirror.
    5. Connect the hose to the CO2 tank by hand and then finally tighten (without excessive force) using the spanner, which is kept with the hose set.
    6. Connect the pipe to the hose
    7. Climb into the mirror cell with the CO2 hose and pipe
    8. Open the CO2 tank while the pipe is opened. (Make sure the pipe is NOT pointing at the mirror!)
    9. Adjust the flow of snowflakes until you have a nice slow flow of snowflakes leaving the pipe
    10. Now point the pipe at the mirror and spray the whole mirror with the snowflakes. Make sure you don't stay too long on one place. Keep the pipe moving. The snowflakes should fall on the mirror and glide down with the dust.
    11. Attempt to clean uniformly by sweeping the plume of snoflakes in a regular way accross the surface, like when painting.
    12. When finished, close the valve on the CO2 tank but do NOT close the valve on the wand.
    13. After you have done the whole mirror wait for 10 minutes and do it again (in the WHT this time can be used to do the Nasmyth flat)
    14. When finished cleaning at WHT, check the control box for the Nasmyth turret mechanism, on the inside of the cube, where you are standing. The switch should be set to ON, otherwise the Nasmyth flat mechanism will not move.
    15. The WHT mirror cover petals on which you have been standing should be pushed upward until you feel resistance from the mechanism. This will ensure they will close fully later.
    16. Tidy away the equipment
    17. Once at least 10 minutes have passed since cleaning finished, you can make the final reflectivity measurements.
    18. Move the telescope to Zenith park
    19. Alternatively take Reflectivity measurements through the ports at senith park.