Helium flexlines


This section deals with the Helium flexlines as they are used in the system to  connect the compressor to the cold head. It will also explain how to connect and disconnect the aeroquip fittings.

In the current system we are using 1/2"  ID flex lines. These are 6m length rubber hoses with a steel braid to protect the rubber. On both ends an  aeroquip connector is fitted. These are made to connect and disconnect without loosing pressure. Each flex line is pressurized with Helium and therefore ready to be connected to the system so there is no need to purge them.

When connecting or disconnecting flexlines you have bear the following things in mind:

Always connect the flexlines to the compressor first, then work towards the coldhead.

When connecting flexlines First connect the return-flexline, then connect the supply-flexline
When disconnecting flexlines First disconnect the supply-flexline, then disconnect the return-felxline

The aeroquip connectors are designed to be connected and disconnected without pressure leakage. Making and breaking the connectors is straight forward. The only thing you should watch for is the O-ring that is fitted on the male-connector. As long as this in place hardly any leakage will occur. 

To connect a connector: tighten the two halves handtight, then use special spanners (Ingrid-cupboard) to tighten.                                               To disconnect a connector, use two special spanners to undo, make sure you undo the two coupling-halves and not the whole connector.


Last Updated 10/04/2002, Michiel van der Hoeven