Alignment of the prisms on the SmallFibres module
Important
to know what the tolerances are on the alignment of these prisms.
The
prisms are 44-44-92 degrees. This is for compensating the telecentrisitie of
the beam coming from the telescope on to the fieldplate. There fore the beam
entering the prisms will have a maximum difference of about ±2
degrees. This means that the accuracy of the alignment of the image exiting
the prism has to be within ±0.5
degree.
Also
the position of the prism to the button is not important for the placing robot
can cope with this.
The
alignment of the rotation in the fibre axis is important. This should be
aligned as accurate as possible.
The
next steps will guide you through the alignment of the gluing jig and the
gluing and aligning of the prisms.
Alignment
of the gluing jig
Fire
a laser on the right height, where the prism will be, to the jig. On the
position of the prism put a flat mirror mounted to the jig.
Align
the laser until the reflected beam I goes back into the laser.
Put
a target under the laser exit, which has its centre 4 degrees down. The
distance of the target centre to the laser exit depends on the distance of the
target/laser to the prism.
Gluing
of the buttons and the prisms
Things
to do before gluing the buttons and the prisms
-
Glue
the magnets in the buttons. Make sure that the magnet is within the button and
that glue isn't spilled on the reference surfaces of the button.
-
Look
for orientation of the prisms. There is a facet off the prism, which indicates
the coated side of the prism. You can also look for the uncoated surface of
the prism. This side has to be glued on the fibre
-
Clean
the fibre ends
Gluing
the buttons and prisms
1)
Mount the button with the fibre on the jig against
the mechanical top at the top.
2)
Align the fibre with target behind the jig.
3)
Put some glue on the fibre end.
4)
Check if the fibre is still at the stop or bring it
to it again and check the alignment.
5)
Bring the prism to the fibre end
6)
Align the prism to the target within its tolerance
band (+- 0.2 degrees)
7)
Lower the prism and check / realign the prism
8)
Bring in the UV-lamps and let the glue harden for
100 seconds
9)
Take out the fibre from the jig and check the
cosmetic quality of the gluing
10)
Continue to the next fibre