ΨPerform a white light flat:
to equalise in piston
Switch on Simplex light,
Iris = -150
Set WFS Integration Time to
25ms or 50ms
Spots have to be bright
enough for the algorithm to converge
Type AutoWhite from Navis prompt
Answer 0 (leave alone) to
the question about Integration Time
The process takes more than
5 min; the X-shifted and Y-shifted lenslet
Note the result: the algorithm gives
satisfactory result if:
There are more than 120
measurements
X and Y gradients are less
than 200 in absolute value
For specialists: If Gradients are high, take
out a global tip (X) or tilt (Y) with AddZernike tip (gradient/100) then the return flat will need adjusting in Grace, and a new
Laser/White light flat
After Laser Flat and White
light Flat, dont forget to restore air flow to >80% in Grace.