"This procedure includes enough checks..."
  • This procedure includes enough checks to serve as “pre-AO run” check list. Not everything would be needed on a normal day after the run has started
  • In GRACE
  • Outside of GRACE
    • Starting up the system
    • Laser flattening the Deformable Mirror
    • White light flat
    • Simplex flat and checking image quality














2
"In GRACE"
  • In GRACE
    • Environment
      • Check if humidity is acceptable (Hygrometer on the table); switch on or off the de-humidifier accordingly, empty its tank
    • Inspection of the light path
      • Take out all mirror covers (DM, OAP1 flat, FSM, Fisba small return flat, OAP2 flat). Place covers in the dedicated plastic rack
      • Check if nothing obviously obscures the light path (forgotten tools, mask). More subtle obscuration will need imaging with WFS to reaveal itself.
      • Check which dichroic is set: 50/50 for observations with OASIS, IR dichroic (Barr) for INGRID/OSCA. From Nov 2005, dichroic changer are controlled from oasisgui or a dedicated PC in the Control Room. There is no physical contact anymore.
      • Check which pinhole is set at the Simplex NCU (need 2mm)
    • Electronics
      • Reboot the FISBA PC every other day (all automatic; simply “Restart”)
      • Check if all VME racks are alive (all green…). This can be tested later on from TopGui or an engineering Gui at Navis.













3
"In the Control Room"
  • In the Control Room
    • Check on the air-conditioning control panel if temperature is stable
    • Run up obssys on taurus, check cameras:
      • AG3, AG7, INGRID, MITLL3 (for OASIS)
    • From NAVIS rightmost console and keyboard
      • Start a whtobs sesion with a telnet to Taurus (obssys, 1, 1). Before telnet don’t forget to xhost + to authorise graphics from Taurus onto Navis.
      • startoasisgui ; if not working try first shutdownoasisserver; startoasisserver; startoasisgui
      • From oasisgui float the bench; move NCU to simplex
      • Check if all mechanisms are in green on USP, Ingrid, OASIS pages (otherwise a corresponding VMS rack will need reboot)













4
"From NAVIS leftmost console and..."
    • From NAVIS leftmost console and keyboard
      • See if TopGui is running, otherwise issue a restart or NaomiRestart and topgui & at the NAVIS prompt. There is a “special case” recipe in the green Naomi Software folder, at the tab Trouble (hand written in the margin)
      • From the Light Path page load the latest DM flat. After a long period of inactivity (> week) try out several previous flats if needed. For Fisba to work fine, interference fringes should be seen on every segment.
      • From the C40 page Start WFS framing, close the SG loop
      • Check if the SG gain is 0.8, otherwise type SG SetGain 0.8
      • Check if the  WFS is ok and Click on Continuous
      • Check noise level on the WFS, visually or with a Grab
      • Check on the NAOMI logbook where was the pickoff to see spots. If the pick-off doesn’t move, INIT+START and INDEX+START from the engineering gui launched from EngGuis – Mechanisms DM
      • Sanity check for the SDSU: try changing to Mode 10 (while the WFS is framing) then switch back to Mode 1

















5
"From oasisgui switch on the..."

    • From oasisgui switch on the simplex lamp
      • Spots should be visible on the WFS window. Otherwise  (and if the pickoff is at the right position, NCU light is on, shutter open) something has been left on the light path!
      • As a sanity check, from TopGui try moving the FSM in X and Y, from Light Path Page: Mid range is for 4096 adc; go to 5192, 3072, you should see the spots move to the corner of a WFS cell if they were centered (1 arcsec = ˝ cell). Then Midrange (not Zero!)
      • Light should fall on AG7 (OASIS+Simplex mirrors); run 0.1 AG7 from Taurus, or check from the simplex:
        IngridAlign.py --camera AG7
      • or mirroralign –camera AG7
      • At this stage if the image looks good enough (fwhm <3 pixels) the checks are done
















6
"Perform a laser flat:"
    • Perform a laser flat: equalise elements tip and tilt
      • Go to the air conditionning console (upper left corner above the TO console, next to the met station) and set Fan speed to 20%
      • From TopGui Engineering Menu launch FISBA
      • Click on One Shot , if nothing happens the FISBA PC needs rebooting
      • If all DM elements are not visible (i.e. Fringes not present on some segments) tweak by hand (if loading the latest flat was not enough)
      • Click on Iterate
      • Save this flat from the Light Path page DM controls
      • All DM elements should show one fringe only. Now they are all parallel but maybe not in the same plane














7
"Perform a white light flat"
    • Perform a white light flat: to equalise in piston
      • Switch on Simplex light, Iris = -150
      • Set WFS Integration Time to 25ms or 50ms
      • Spots have to be bright enough for the algorithm to converge
      • From the General Scripts window do an Autocentre
      • Check focus; the spots separation has to be close to 8 pixels. Adjust the WFS fore-optics focus from the EPICS guis
      • Type AutoWhite from Navis prompt
      • Answer 0 (leave alone) to the question about Integration Time
      • The process takes more than 5 min; the X-shifted and Y-shifted lenslet
      • Note the result: the algorithm gives satisfactory result if:
        • There are more than 120 measurements
        • X and Y gradients are less than 200 in absolute value
      • For specialists: If Gradients are high, take out a global tip (X) or tilt (Y) with AddZernike tip –(gradient/100) then the return flat will need adjusting in Grace, and a new Laser/White light flat
      • After Laser Flat and White light Flat, don’t forget to restore air flow to >80% in Grace.














8
"Perform a simplex:"
    • Perform a simplex: to optimise image quality
      • Run the simplex (with AG7 visible light or Ingrid for infrared)
      • Set WFS Integration Time to 25ms or 50ms
      • To simplex with Ingrid, take out OASIS and Simplex mirrors (from USP tab on oasisgui),otherwise (simplexing with AG7) you need both mirrors in.
      • Adjust the Simplex Iris so that the maximum intensity measured is between 10000 and 20000
      • Click on Simplex
      • Monitor the Merit Function (MF): image is considered good when MF < -0.015 (with AG7) or MF < -0.03 (with Ingrid)
      • Usually image starts improving after 800-1000 cycles
      • If after 2000 cycles MF doesn’t improve, stop simplex then start again: the algorithm may be stuck in a local minimum
      • When the MF is satisfactory, stop simplex
















9
"Check image quality from the..."
    • Check image quality from the observing system
      • run AG7 0.05
      • Display the .fits file, use imexam j and k commands
      • Image is good when its fwhm is between 2 and 3 pixels



    • Complete description of the NAOMI checks:
    • http://www.ing.iac.es/~crb/wht/aorecipe1.html













10
"Derotator,"
    • Derotator, NCU, Flat, DM. back














11
"Mirror covers rack."
    • Mirror covers rack. back














12
"NCU and pinholes,"
    • NCU and pinholes, DM. back














13
"Fisba PC (in electronics..."
    • Fisba PC (in electronics room). back














14
"oasisgui (showing here usp..."
    • oasisgui (showing here usp mechanisms).  back














15
"oasisgui (showing here ingrid..."
    • oasisgui (showing here ingrid mechanisms).  back














16
"oasisgui (showing here oasis..."
    • oasisgui (showing here oasis mechanisms).  back














17
"Tools menu in TopGui."
    • Tools menu in TopGui.  back














18
"Light Path page in TopGui"
    • Light Path page in TopGui.  back














19
"C40 page in TopGui."
    • C40 page in TopGui.  back














20
"Start framing."
    • Start framing.  back














21
"Close Strain Gauge loop."
    • Close Strain Gauge loop.  back














22
"SDSU (WFS)"
    • SDSU (WFS) page.  back














23
"Shack Hartmann Spots on WFS"
    • Shack Hartmann Spots on WFS.  back














24
"How to check the FSM..."
    • How to check the FSM is alive. If the spots don’t move, suspect a fault or simply go into GRACE to turn it on! (along with the FSM filters, in the first rack from the left in the electronics room)     back














25
"Simplex gui."
    • Simplex gui.  back














26
"Simplex image."
    • Simplex image.  back














27
"Engineering gui’s menu."
    • Engineering gui’s menu.  back














28
"Fisba gui."
    • Fisba gui.  back














29
"Fisba gui."
    • Fisba gui.  back














30
"Mirror control page."
    • Mirror control page.  back














31
"Scripts gui."
    • Scripts gui.  back














32
"Autocentre."
    • Autocentre. Press Enter to make window disappear.  back














33
"WFS Alignment."
    • WFS Alignment.  back














34
"Adjust Fore-optics focus from mechanisms..."
    • Adjust Fore-optics focus from mechanisms engineering gui.  back














35
"Result of AutoWhite."
    • Result of AutoWhite.  back














36
"Simplex."
    • Simplex.  back














37
"Merit Function of the simplex..."
    • Merit Function of the simplex algorithm.  back














38
"Result of the simplex algorithm"
    • Result of the simplex algorithm.  back














39
"Checking image quality with imexam"
    • Checking image quality with imexam.  back