1. Testing AUTOFIB & Westinghouse TV.
4. To test WYFFOS CCD These checks should be completed along with the telescope checks and assume that ICL is up and running and configured for AF2/Wyfoss. Also the AF2/WYFFOS VME and AF2 power supplies are up and running in GHRIL. (see engineering notes for details on how to boot the Micro’s).The AF2/WYFFOS observing system consists of the following sub_systems
WYFFOS (location GHRIL).
WYFFOS VME. (location GHRIL)
Atmospheric Dispersion Corrector. ADC. (location prime focus)
Prime Focus Turntable. PFTT. (location prime focus)
CCD Autoguider. (location AF2 robot at prime focus)
WYFFOS
CCD. (location GHRIL)
1.7 At ICL prompt > AFPARKR. This parks the robot to the edge of the fibre field but the power remains on.
1.8 At ICL prompt > AFMOVP 150000. This moves the off-axis probe. NB the robot must be parked (afparkr) to complete this operation.
1.9 At ICL prompt > AFMOVP 0. This moves the probe to its park position.
1.10 At TV panel ensure that the gain is off then turn on both Prime and Cass systems. The cass system is turned on only to maintain sync.
1.11 At ICL prompt >AFLIGHTS ON. This turns the inspection lights on inside the AF2 robot. NB this would damage the TV if the TV was already on and the gain was high.
1.12 Ensure all domes lights are off. If they are not then they have been overridden. Remove override with key switch before proceeding.
1.13 Now the dome is dark and AF2 inspection lights are ON. Turn on the gain of the prime camera and increase until you see the guide fibres. They are fibre bundles with a fibre in the middle.
1.14 Turn off camera system ensure the gain is fully down and off.
1.15 At ICL prompt >AFLIGHTS OFF. This turns the inspection lights off.
1.16 At ICL
prompt> AFSLEEP. This safely parks robot out of the field
and turns the power off.
This is not a detailed check but proves communications. Moving other things like the grating can disturb the observers set-up
2.1 At ICL prompt> WFILTA IN. At ICL prompt> WFILTA OUT. This checks filter slide A
2.2 At ICL prompt> WFILTB IN. At ICL prompt> WFILTB OUT. This checks filter slide B
2.3 At ICL prompt> WCAL 1 ON. This turns the calibration lamp on. Check it changes state on the WYFFOS mimic.
2.4 At ICL
prompt> WCAL 1 OFF. This turn calibration lamp off.
3. ADC test. Use the wyffos mimic to display corrector element positions.
3.1 At ICL prompt> ADC_AUTO DISABLE. This allow you to move the elements manually
3.2 At ICL prompt> ADC 90
3.3 At ICL prompt> ADC -90
3.4 At
ICL prompt> ADC_AUTO ENABLE
(assumes DMS and sprac real time display is running)
4.1 Turn lights off in the dome.
4.2 Ensure Nasmyth Flat is stowed for GHRIL.
4.3 At ICL prompt> SETUP WYFFOS Default setup is usually OK
4.4 At ICL prompt> BIAS WYFFOS (Check bias frame on SRTD)
4.5 At ICL prompt> WCAL 1 ON
4.6 At ICL prompt> GLANCE WYFFOS 200
4.7 At ICL prompt> WCAL 1 OFF
NB. If you have to re-boot the autoguider the latest version is currently on a floppy disk
6. Re-boot required (If system is up and running go to 5.15)
6.1. Reset Autoguider in rack. Quickly return to the Autoguider terminal
6.1.2 Press >CTRL C to abort task.
6.1.3 >Insite reload. This forces the system to boot from the floppy.
6.1.4 When promted select >AF2
6.1.5 Check to see if CCD is cooling or cold if its not type >Start-up
6.1.6 >TV ON check sensitivity to light. Toggle between fields with FM1 and FM2. Do not use the Field command the system hangs if there is no star. When