TEK3 Parameters
Device Identification
Chip name Tek3
Year of manufacture 1994
Serial number 1578BR04/01
First light on La Palma May 1994
Description Thinned AR coated
Operational Characteristics
Speed |
Slow |
Standard |
Quick |
Turbo |
Nonastro |
Bias(ADU) |
|
810 |
877 |
||
Gain(e/ADU) |
0.7 |
1.39 |
|||
Noise(e) |
4.7 |
6.1 |
|||
Linear to(ADU) |
60k |
60k |
|||
Readout time(s) |
166 |
140 |
Vertical CTE 0.99999
Horizontal CTE 0.99999
Physical Characteristics
Pixel Scale (235 camera) 0.70 arcsec/pixel
(500 camera) 0.33 arcsec/pixel
X Pixel size 24 microns
Y Pixel size 24 microns
X size in pixels of digitised area 1124
Y size in pixels of digitised area 1124
X size of useful imaging area 1024
Y size of useful imaging area 1024
X start of useful imaging area 50
Ystart of useful imaging area 0
LN2 capacity of cryostat 2.5 litres
Cryostat window thickness 4 mm
Distance from window to CCD 11.1 mm
Cryostat window size 40mm x 40mm
Operational Parameters
Flooding type none
Operating temperature 173K
Prefered amplifier A
Network name CCD1
Anti-blooming available no
Measured characteristics
Quantum Efficiency
Dark current 9 e/hour
Full well bloom limit ?
Chip flatness ? micron ? bow
Cosmetics ?
This page last updated : 17th June 1998