Device Identification
Chip name Tek2
Year of manufacture 1993
Serial number 1396CR02-01
First light on La Palma November 1993
Description Thinned AR coated
Operational Characteristics
| Speed | Slow | Standard | Quick | Turbo | Nonastro |
| Bias(ADU) | 996 | 1088 | 1231 | 849 | 379 |
| Gain(e/ADU) | 1.07 | 1.19 | 1.62 | 2.45 | 4.56 |
| Noise(e) | 5.22 | 4.99 | 5.77 | 10.26 | 22.6 |
| Linear to(ADU) | 60k | 60k | 60k | 50k | 15k |
| Readout time(s) | 164 | 112 | 73 | 58 | 40 |
Vertical CTE 0.99999
Horizontal CTE 0.99999
Physical Characteristics
Pixel Scale (ISIS red) 0.36 arcsec/pixel
Field of view (?) ?
X Pixel size 24 microns
Y Pixel size 24 microns
X size in pixels of digitised area 1124
Y size in pixels of digitised area 1124
X size of useful imaging area 1024
Y size of useful imaging area 1024
X start of useful imaging area 50
Ystart of useful imaging area 0
LN2 capacity of cryostat 2.5 litres
Cryostat window thickness 4 mm
Distance from window to CCD 11.5 mm measured optical path length
Cryostat window size 40 mm x 40 mm
Operational Parameters
Flooding type none
Operating temperature 173k
Preferred amplifier A
Network name CCD6
Anti-blooming available no
Measured characteristics
Dark current 6 e/hour