Chip name SITe2
Year of manufacture 1995
Serial number 4321ADR01-01
First light on La Palma December 1997
Description Photo of Site chip in cryostat
Operational Characteristics
|
|
|
|
|
|
|
|
| Standard |
|
|
|
|
|
|
| Quick |
|
|
|
|
|
|
| Turbo |
|
|
|
|
|
|
| 2x2 Binned |
|
|
|
|
|
|
Vertical CTE 0.9999993
Horizontal CTE
0.999999
Physical Characteristics
Pixel Scale (JKT A&G) 0.33 arc/pixel
Field of view (JKT A&G)
10 x 10 arcmin (unvignetted)
11 x 11 arcmin (vignetted,70% light loss)
X Pixel size 24 microns
Y Pixel size 24 microns
X size in pixels of digitised area 2148
Y size in pixels of digitised area 2148
X size of useful imaging area 2048
Y size of useful imaging area 2048
X start of useful imaging area 20
Ystart of useful imaging area 1
LN2 capacity of cryostat 2.5 litres
Cryostat window thickness 4 mm
Distance from window to CCD 10.3 mm
Cryostat window size
80mm diameter
Operational Parameters
Flooding type none
Operating temperature 156K
Preferred amplifier A, (B is possible backup but noisy >20e)
Network name CCD4
Anti-blooming available
no
Measured characteristics
Quantum Efficiency
Dark current 1 e/hour
Full well bloom limit 130000 e
Chip flatness 230 micron convex bow
Cosmetics Two blocked column clusters close to edges of chip