1. Switch off controller and connect the SCSI disk rack to the back.
Note: Any Forth SCSI rack with an INSITE drive can be used - there
is one dedicated for this purpose, but the DMS, TV and Autoguider racks
can be used just as well.
2. Switch on SCSI rack first
3. Then switch on CCD controller. If the controller locks up as it loads the RAM disk the reboot the controller, pressing control C as soon as you see the ImageForth banner - ie before the RAM disk is loaded.
4. Put the 3.25" floppy from the front of this folder into the Insite drive - as backup copy is available from the WHT CCD instrument specialist.
5. type INSITE 1 LIST - this displays the instructions and locations of the backup for each CCD. Note down the start block for your detector.
6. Using the number specified type:
no RESTORE-EEPROM
Note since Version 3.0 of the application - RESTORE-EEPROM has been
available without loading it from the disk.
7. Reset the controller, and check the code gets recomipled correctly - and check SEE comes up with the correct detector name, and the network name is correct.
8. Switch off CCD controller and remove SCSI rack, switch CCD controller
back on, Reset the DICI on the DMS.
1. Turn off controller and connect SCSI disk rack. Turn on SCSI rack and then CCD controller
2. At terminal - check the controller has booted up and the RAM disk is not corrupt.
3. Type INSITE 1 LIST and find a suitable place for the blocks - there is no need to overwrite the old copy of the RAM disk - you may wish to return to it.
4. Type: INSITE 1 DRIVE no. SAVE-EEPROM Where "no." is the block number selected above.
5. Reboot controller and check all is functioning as before.
6. Power off CCD controller then power off and disconnect the SCSI rack and power up the CCD controller again.
7. Reset the DICI on the DMS.
OPEN Open shutter
CLOSE Close shutter
UNJAM Unjam shutter
?OPEN Check shutter status -1 for open, 0 for closed
SEE Show general status
1 TELE Display telemetry for relevant head.
2 TELE
WHO? Display which CCD you are talking to.
T-SHOW Display temperature of the two heads
HOR Start clocking horizontal clocks continuously
VER Start clocking vertical clocks continuously
PIX Send test data down the fibre continuously - FOX data & FIFO
leds should light
NON Disable the above three commands
REBOOT Reboot controller - reset DICI afterwards
RAM-DISK 1 LIST etc check for corrupt RAM disk
ddmmyy SET-DATE Set date to dd/mm/yy
hhmmss SET-TIME Set CCDC time to hh:mm:ss