Spectral Response Curve for the UES CCD

The pixel size is 22 microns square and the surface has been coated to enhance the blue response. The rather longish delay in providing this large format device has been due to the anomolous dark current we measured (now almost 2 years ago! ) several hundred electrons per hour. As this is a 2-edge buttable detector, the packaging has meant it was difficult to secure the device to the cold block and we hence assumed that this was the cause of the high dark current. I have recently been involved in some work to 'glue' the chip down onto a piece of Kovar and this has resulted in a package which can be handled more easily. The Kovar is secured by 4 retaining screws and we are confident that this has now ensured that the CCD is making appropriate contact with the block.. Measurements are under way. The goal we are seeking is a dark current of around 1-2 e/hour and consistent with what we measure on other devices of similar type. We hope to have it on the telescope by Christmas.

Spectral Response of the UES, S/N=A3204-18-11 CCD
Response Curve