A CCD detector for multiple object spectroscopy at the United Kingdom Schmidt Telescope.
A. P. Oates
Department of Physics, Science Laboratories, Durham University, South Road. Durham. DH1. 3LE. U. K.
ABSTRACT
At UKST (United Kingdom Schmidt Telescope, a unit of
the Royal Observatory, Edinburgh. U.K. now operated by the
Anglo-Australian Observatory, Epping. N.S.W. Australia)
a CCD camera has been installed to operate with
the FLAIR (Fibre Linked Array Image Re-formatter) fibre
system (see Watson). The
standard EEV P8603 CCD used in the camera has a poor short
wavelength response and so all the CCDs used at UKST are
surface coated with a fluorescent dye to partially overcome this
problem. In 1988 it was decided that the system's
response around
could be improved further by replacing
the FLAIR fibres with a set of fibres offering superior
transmission properties at this wavelength. The
introduction of these (larger core diameter) fibres would have meant,
however, accepting a reduction in signal-to-noise as the fibres
illuminate more pixels on the CCD. The CCD sequencing was therefore
modified to enable us to on-chip bin pixels (across the dispersion
direction) to alleviate the problem of reduced signal-to-noise
which would have otherwise occured. Most
recently, modifications have been made to the CCD system to provide
detection capability for approximately twice the previous number of
objects. This has been achieved by appending a second CCD
detector and correlated double sample(CDS) processor to the
existing sequencer.
Both CCDs are operated via a signal controller which can route clocks
and video between the detectors and sequencing electronics.
Reduction of galaxy data, obtained during May/June 1988 and
August 1989 are to be published in a forthcoming paper.
These data have shown that FLAIR, combined with
a low noise detector, in both single and dual CCD mode, is easily
capable of obtaining cross-correlation red-shifts in the blue
with a high success rate.
1. CCD SYSTEM
The CCD camera being used at UKST was developed
by Waltham and initially used with FOS 1 (the Faint
Object Spectrograph
) on the U.K. Isaac
Newton telescope. The system is controlled by a DEC PDP11/23
which, at the time the system was built was a fairly standard
mini-computer to use for this purpose. The system consists
of the components illustrated in figure 1 and described below.
1.1 CCD front-end
The CCD camera utilises an EEV P8603B image
sensor housed in a standard liquid Nitrogen filled dewar. The
cryogen capacity of the dewar results in a hold-time of
hours (this is due to the axial fill tube arrangement
and to the fact that we use the dewar horizontally) and maintains
the temperature of the CCD at 150K. This temperature was found
to be the best compromise in terms of optimising read-out noise,
dark current and charge transfer efficiency. A custom PCB at the
front of the dewar supports the 30-pin, socketed CCD and the
components of a Wheatstone bridge circuit, one arm of which
contains a platinum resistance thermometer
(T.C. =
;
at 273K) which forms
part of the temperature servo-circuitry. Bias supplies,
(the CCD output) and the horizontal and vertical
clocks are all conveyed to the CCD through 9-pin vacuum feed-through
connectors; all bias supplies are de-coupled close to the CCD.
Along with low read-out noise,
blue responsivity was the only other main requirement of the
detector (assuming good horizontal and vertical charge transfer,
above average numbers of cosmetic defects can be tolerated, though
in devices used so far, these have been few in number). The CCD
was coated at ESO
with an organic polymer to enhance
its short-wavelength sensitivity. The coating produces an improvement
in quantum efficiency(QE) in the blue of, typically, 15%. Although
better detection response has been obtained at shorter wavelengths,
the system's response in the
wavelength range has
been improved still further with the introduction of a new
blue-transmitting Polymicro fibre feed; the combination of
this with on-chip pixel binning across the dispersion direction has
significantly improved the system's signal-to-noise ( S/N)
at short wavelengths.
1.2 CDS processor
The CCD output (typically ) is taken, via
a constant current source, to the input of a very low-noise
operational amplifier within the CDS and provides
a first stage signal gain of 37. A constant current load was
chosen in preference to a resisitive load as the former resulted
in better gain from the on-chip FET amplifier. Analogue switches,
driven from the sequencing electronics, enable the following control
during pixel read-out:-
the CCD output node is RESET,
control clock RAMP-UP is enabled for
to sample the output node voltage prior to charge transfer,
a 3-phase horizontal shift moves charge from 1
pixel in the serial register onto the output node in
,
control clock RAMP-DOWN is enabled for a further
to sample the RESET+VIDEO signal.
The difference between the first and second sampling voltages
is a measure of the pixel charge; the dual slope integrator also
provides a further signal gain of 5. The video is then digitised
by a 16-bit successive approximation
analogue-to-digital converter which was selected for its fast
conversion time (16-bits in ) and good linearity.
Pixel read-out is completed in
and this is
the rate at which digitised data is available from the A/D
converter during a one-line-sequence period (
).
In 3
2 pixel binning mode, the whole CCD can be
read-out in 17 seconds.
1.3 Driver electronics
A CCD sequencer generates low voltage D.C.
which are conveyed to precision voltage references within the
CDS processor for the CCD bias electrodes. Table one
lists the current settings of the CCD bias voltages and clock read-out
parameters.
CCD control is provided by the sequencer
through use of control EPROMS. These are accessed by the
PDP11 which commands a one-line sequence (400 pixels)
of the CCD whereby a pixel is read-out, digitised to 16 bits
and stored in memory (within CAMAC) for subsequent
retrieval by the PDP11. This procedure is repeated upto 578
times to enable a full-frame read-out of the EEV CCD. The
observer may select vertical binning sequences of 1, 2, 5 or 10
pixels to ensure that dispersed spectra are correctly sampled
when using the large diameter Polymicro fibres (spectral
resolution /pixel with the
/mm grating).
Instructions to achieve the control described above are
issued by a FORTRAN program running under RT11. This
menu-based program supervises all CCD operations and enables the
observer to construct exposure sequences to suit particular
requirements. A simple command language is available for this
purpose so that custom exposure sequences can be constructed and
executed automatically by the sequencer. Provision also exists
to store these command sequences on disc for future use. Data
read from the CCD is stored on disk and displayed as an 8-bit grey
scale on monitors in the dome and at a remote location. Here
the observer can more conveniently control the operation of the
camera using a vt100 type terminal, away from any extremes in
temperature present in the Schmidt dome.
Data obtained from earlier runs may be inspected whilst the next
exposure is in progress, during which time the dome has to be
kept in darkness.
2. PERFORMANCE
The following is a list of observations which have been
noted during the general operation of the system over the last
3 years:-
Operating a system in a dry enviroment can lead
to problems with static charge. Thick earth returns were installed
from the dewar, pre-amp electronics and optical table to the
star earth within the CCD sequencer. The optical table
provides a stable base for the CCD camera and spectrograph and
during observations is mechanically and electrically isolated from
the dome floor.
Only one problem has been encountered with
the passive components on the dewar PCB. A decoupling capacitor
failed and resulted in spurious operation of the CCD when at
operating temperature. The temperature of these components has
been measured and found to be 180K when the CCD was at its
operating temperature of 150K; it is unknown what long-term
effects such temperatures have on these components.
No problems have been encountered with the
polymer coatings of our CCDs in terms of adhesion to the CCD surface.
No adverse effects have been found either with
the CCD or with the polymer coating when cooling the CCD from room
temperature to its operating temperature in
45minutes
(i.e.
3K/minute).
Because of the modest aperture of the telescope,
the read-out noise of the detector is the limiting factor when
attempting to obtain data with good S/N. By exposing the CCD
twice to 10 different levels of illumination, it was possible to
obtain the mean count by averaging over two regions on each of the
two frames and de-biasing from the 11 pixel underscan strip;
variance was obtained after subtraction of two identically
exposed frames and the result plotted. From this data the
system gain (G) was found to be-
and the CCD read-out noise (
) is-
In addition to the well known temperature
dependence of dark current, there is also a dependence of read-out
noise charge transfer efficiency(CTE) and QE with temperature
.
In the present case it was found that the CCD could be operated
anywhere in the temperature range-
without any significant change in the read-out noise .
The horizontal and vertical CTE of this device
appears to be very good. Some tuning
of the horizontal clocks and adjustment of the voltage on the
CCD output gate (
) was made in the initial stages
and this resulted in, at most, only 1 pixel smearing of cosmic
ray events. The 4-pixel overscan was also investigated for residual
charge on a low light level flat-field; the mean level of charge
was found to have fallen back to the bias level after 1 pixel.
The temperature dependence of CTE was not investigated.
The 5/10/10BC CCD does not exhibit any measurable
dark current in exposures of 1 hour; this being the maximum
exposure time used per frame, if the effects of cosmic ray events
(CREs) are not to contaminate the data too severely.
Our integration times are limited to around 3000s
to prevent cosmic rays contaminating our data too severely. This
device type usually produces
2 events/cm
/minute but this
was exceeded in our early data by a factor of 3-4. The source of
the excess CREs was investigated
and found to be caused by the glass window at the front of the
CCD dewar. This window is made of Schott glass some of which,
it appears, emit low-level X-rays which can be detected by the CCD.
The solution was to replace the original window with one made of
fused-silica
.
3. RECENT DEVELOPMENTS
During 1988 two proposals were made in an attempt to
up-grade the system. The first related to the problem
of the system's lack of sensitivity at the blue end of the spectrum
(around ) and the second related to how we could modify
the front-end so that we could increase the number of objects
observed during a given exposure.
3.1 Improving sensitivity
To try and overcome
the problem of the system's poor response at short wavelengths,
a new fibre feed was installed whose transmission properties in
the blue were superior to the existing FLAIR m
fibres. To ensure that a comparable S/N could be maintained
with the new Polymicro fibres, pixel binning across the
dispersion direction was introduced. The new fibres have a core
diameter of 100
m, the optics introduce a magnification of
1.7 hence it was thought at first that a 7+3
(170
m/22
m & 66
m/22
m; the EEV
pixel size is 22
) pixel binning sequence
should be introduced; the inter-fibre gap on the CCD is 3 pixels.
It was, however, impossible to ensure that the output of the fibres
would keep exactly in step with pixels binned in this fashion along
the entire 400 pixel row (it was felt that the fibre output would
get out-of-step with the pixel pitch). A less
critical 3-pixel binning scheme was therefore introduced which
resulted in a fibre coverage of 2 binned pixels (6 CCD pixels)
and an inter-fibre gap of 1 binned pixel (3 CCD pixels). Initially,
single pixel read-out was accomplished by having charge on the
output node sampled as
and
of the serial clocks
were changing state. The transition times for the clocks, and
hence the pixel integration time was therefore relatively
short,
, see table one. As any 1/f component
of read-out noise (
) is dependent on this
time, it is kept as short as practicable. However, to
compare
in un-binned and
binned CCD frames,
the sequencing was changed so that a complete shift of
the serial clocks occured when reading out a single pixel;
the pixel time (
) now being
.
The read-out noise was measured again and found to be comparable
with that obtained using the old pixel read-out sequence. From
here it was a (relatively) simple matter to increase the number
of complete serial shifts during
a pixel read-out, to enable any number of CCD pixels to be
binned. Although some clocking flexibility is available
from the sequencer, a downloadable sequencer would have enabled
modifications to be made to the clocking patterns without re-course
to blowing sets of EPROMS
each time a new sequence pattern was required. Fortunately this
operation was only carried out 2 or 3 times before we were
satisfied that the sequence pattern was correct. Again the
read-out noise of the CCD was measured for the 2 schemes which
were tried, 3 and 5 pixel binning, there being no change in
read-out noise compared with that from single pixel read-out.
3.2 Improving detection capability
The second proposal was that we could, at reasonable cost, append a second detector head to the existing sequencer, permitting us to increase (from 35 to 71) the number of objects observed during a given exposure. The primary motivation for this was that it was the most expedient way we could realistically commence work on a large survey of galaxy redshifts.
3.2.1 PC CCD system
To achieve the above we required a second detector, this
time to be dye coated by John Barton at the AAO, a new cryostat,
pre-amp and electronics designed to perform the switching operation
between the 2 CCDs. The system would be tested in Durham using a
laboratory CCD camera which had been built during 1988/89
using front-end CCD hardware similar to that in Australia.
It was decided to develop the new system interfaced to a
PC running GEM Windows 3.01. The 20Mhz 80386 PC,
supplied by the U.K. firm Elonex, also included 1Mbyte
of RAM, an 80387, 20MHz co-processor, a 44Mbyte hard disk,
DOS 3.3, VGA and serial mouse. To manage the operation of
the camera, a new application was written in 'C' and
80x86 assembler, the opportunity being taken to add new features
to the system software. Extra hardware (figure 2) was also
required in the form of-
a 48-bit I/O card to communicate with the
CCD sequencer, permitting 16-bit control, 3-line interrupt
and 16-bit data capability through a new CCD interface. The
I/O card was supplied by Amplicon in the U.K.
an opto-isolated CCD interface, approximating
the function of the CAMAC crate in the system at UKST.
Designed and made in Durham, and
new image display hardware providing
pixel resolution at 8-bits per pixel grey
scale or false colour, hardware pan and scroll, image processing,
zooming etc., significantly better than the image display at
UKST. This was supplied by the Canadian based firm Matrox.
3.2.2 Implementation
It would obviously have been extremely useful to have the dual mode of operation managed from the CCD program, however there was both insufficient time to modify the FORTRAN code on the PDP11 to cope with the dual operation of the system and there was a problem in that the system up to this modification was operating at its limit due to the (now) somewhat out-dated computer hardware. Switching between the two CCDs would therefore be accomplished manually, either from the dome or from the remote location. The additional electronics providing control of the two CCD camera heads being housed in the box containing the new CDS processor and CCD bias supplies. Connections to both the old and new electronics boxes were changed to military specification devices and each CDS box made identical so that it would be a simple matter to revert to the old system configuration should this be required.
Two problems were identified,
(1) how charge on one CCD could be maintained whilst the other was being read-out after charge integration and (2) as horizontal and vertical clocks from the sequencer are supplied to both CCDs, but can only be optimised for one device, it was unclear whether this would compromise the performance of one CCD or the other.
3.2.3 Installation
A simple solution to the first problem was found during discussions with John Barton at the AAO. Charge could be maintained on the CCD not being read-out by supplying both devices at all times with high impedance quiescent voltages (of -9 and -7 volts, similar to those available from the vertical clock driver card during charge read out). The low impedance vertical clocks being easily capable of supplying enough drive to overcome the quiescent state of the CCD electrodes during charge read out. So in addition to the switching function, the new electronics also provides high quality buffered supplies to both CCDs vertical transfer electrodes. During an exposure each CCD simultaneously receives dispersed light from upto 35(CCD I) & 36(CCD II) objects, whilst maintaining the quiescent voltages on the vertical electrodes of both CCDs, sequencing waveforms are applied first to one CCD and then switched to the second CCD and charge read-out from this device.
The system was put together and tested (as far as possible) here in Durham during the Summer of 1989, with the laboratory PC CCD camera acting as the first detector (see figure 3). Due to shortage of time (we were due to observe with the completed system in August/September 1989) we had to wait until the camera was installed at UKST in August, before the second of the problems, mentioned above, could be resolved. In the event it turned out that only slight adjustment to the horizontal transfer clocks was needed to get the new CCD working straight away. There did not appear to be any detrimental change in the performance of the old CCD when running with the new device attached to the sequencer clock drives.
Both CCDs were tested for changes in read-out noise ()
and CTE when being driven separately or together from the sequencer.
The original device showed no adverse changes in either of these
parameters when the new CCD was running from the sequencer at the
same time. The new device, appeared to be somewhat sensitive to
the upper level of
in terms of
; there being
a factor of 2 improvement in
(from
to
) for
a 0.1 volt downward change in
. Small changes in
did
not appear to have any significant effect on
. The new
CCD also exhibited good CTE. There did not appear to be any change
in these parameters when both CCDs were operating from the
sequencer concurrently.
Presently, investigations are in hand to determine whether we could duplicate the PC based CCD system at UKST. The plan has much to commend itself in terms of:-
the PC system would completely replace the
ageing PDP11, RL02 disk drive, CAMAC, and tape drive. It is
faster, cheaper, more reliable and much easier to obtain spares if
these should be required,
better image display hardware,
pixels by 8 bits/pixel, full frame buffering, image filtering,
panning, zooming etc.,
more flexible system control, the second CCD
camera can be controlled from the CCD program. Other features,
such as spectrum extraction and manipulation which already exist
on the system in Durham, could be added as required,
interaction with the program is via a front-end
graphical user interface; the system is therefore very easy to
control. There is also a command processor and compiler (for
the die-hards) to enable custom exposure sequences to be
constructed.
The existing UKST sequencer would plug directly into an interface card which has already been built and tested, hence there would be no changes required to the front-end CCD hardware. The only problem which remains to be solved is that of data archival. Two solutions to this would be:-
(1) to Ethernet data directly to the UKST
micro-Vax and, after observing, TRANSFER all CCD data frames to
the AAT VAX cluster for data reduction (there is currently no
Ethernet link from the UKST micro-Vax to the
AAT VAX). Plans are already in hand
in Durham to run thin Ethernet around the Physics Dept.
with the purpose of enabling user PCs to be interfaced to
our VAX cluster.
This facility is eagerly awaited as, at the moment, the only way
to transfer data to our VAX is via Kermit. This takes
minutes (at 4800 baud) per full CCD frame and was one
of the main reasons for installing some spectrum extraction routines
into the CCD program. This enabled us to select the parts
of the data frame which were to be transferred to the VAX,
speeding the transfer time by a factor of
for the work
currently in progress.
(2) install a PC tape drive which are widely available, can hold large quantities of data and would be easily transportable, though it would require the observer's home institution to support such hardware on their main-frame computer.
4. CONCLUSIONS
Reduction of galaxy data (to be published in a forthcoming
paper, see also Watson), obtained using the system
in both single and dual CCD mode
has shown that we can now obtain cross-correlation red-shifts with a
high success rate. It was clear from this data that CaII H & K
absorption lines, absent from previous un-binned CCD exposures using
the original FLAIR
fibres, were being detected
in single exposures of 3000 seconds (to obtain the desired S/N
(
), 3 or more exposures are made on the same field which
are subsequently added together during the data reduction process).
Figure 4 is a representitive spectrum taken from the June 1988 run
using FLAIR with the new Polymicro fibres and on-chip pixel
binning. This spectrum illustrates the data quality obtained for an
E/S0 spiral galaxy after adding 5, 3000 second exposures; the detection
of the Ca H & K lines in the blue can be clearly seen. The data were
inspected after each 3000s exposure using some rudimentary facilities
available during execution of the CCD program. This gives the observer
some indication of how the S/N of individual galaxy spectra is
progressing as frames are added together. There are
no facilities on the PDP11 to remove CREs which become progressively
more of a problem as frames are added. Data are archived to a
FITS format tape when the RL02 hard disc becomes full, or at
the end of the night's observing, enabling data reduction to be
subsequently undertaken on one of the U.K. S.E.R.C. Starlink
VAX nodes.
5. ACKNOWLEDGEMENTS
The author would like to thank Tom Shanks (Durham) for providing the initial impetus for the project and Fred Watson (R.O.E., Edinburgh) for the work he has done on the UKST FLAIR system. Dick Fong (Durham) provided financial support during the latter stage of the project, thanks go to him and to Mike Breare and John Webster (both at Durham) for their continued support (technical & otherwise) during the project. I would also like to express my gratitude to John Barton at the AAO, Epping for many useful discussions relating to the operational performance of CCDs in general and to the particular problems encountered with our own devices. This work was supported by the UK SERC to which due thanks are also extended.
6 REFERENCES
1 Watson F.G., Proc. S.P.I.E., 627, 787. 1986.
2 Cullum M., Deiries S., D'Odorico S. & Rei R.,
Astron. & Astrophys., 153, L1. 1985.
3 Deiries S., The optimisation of the use of CCD detectors in astronomy. p. 73, June 17-19. ESO. Ed. Baluteau J.-P & D'Odorico S. Pub. E.S.O. 1986.
4 Waltham N.R., The development of a detector system for faint object spectroscopy on the Isaac Newton telescope, Ph.D. Thesis. University of Durham. 1987.
5 Breare J.M., Cox G.C., Ellis R.S., Martin G.P., Parry I.R., Purvis A., Waltham N.R., Webster J., Fosbury R.A.E., Gellatly D.W., Jorden P.R., Lowne C.M., Lupton W.F., Powell J.R., Thorne D.J., van Breda I.G., Worswick S.P. & Wynne C.G., Mon. Not. R. astr. Soc., 227, 909. 1987.
6 Thorne D.J., Jorden P., Waltham N.R. & van Breda I.G., Proc. S.P.I.E., 627, 530. 1986.
7 Mackay C.D., Charge-coupled devices in astronomy. Ann. Rev. Astron. Astrophys., 24, 255. 1986.
8 Watson F.G., The FLAIR wide-field multi-object spectroscopy system , these proceedings.
Table one - Operating parameters for the 5/10/10BC CCD
V refers to a bias electrode voltage; refers to one of the
horizontal, reset or vertical CCD clocks;
refers to the
amount of clock overlap within a 3-phase group; u,l refer to upper
and lower clock levels & r,f refer to clock rise and fall times.