Device W7-4 CCD has been measured at various temperatures,
in imaging AND diode mode. We now use a stable Cadmium source (with
a strong line at 228nm) for flooding rather than Mercury.
This was prompted by a mail message from Mike stating that Cadmium
was to be prefered to Mercury as it produced a better more stable response.
We measure the response at room temperature using either-
- a stable light source and a set of filters with 5nm band-pass, from
330nm to 1060nm, in what is termed Diode Mode. This mode connects ALL
electrodes on the chip together except RD and OD which are paralleled and
returned via a sensitive pico-ammeter to substrate. We generally measure
currents in the range 2-3nA in the UV to 200nA in the mid-band,
- By using a monochromator with a stable source, measuring about every 5nm
across the band 400nm to 1000nm, this feeds the same pico-ammeter and works
fine. A second order filter goes in at 700nm, the light path is direct,
i.e. there is no integrating sphere.
- We have used a phase sensitive detector to measure the signal
BUT we can not get the large chips to work at the lowest chopping frequency
we can use (10Hz). By large chips I mean the thin Tek1024 and the Loral 2kx2k.
Going to somewhat smaller (EEV) devices enables us to obtain a signal but
this signal depends on the chopping frequency - the higher the chopping
frequency the lower the measured signal. At 10Hz the Tek and Loral CCDs
dont produce any signal at all. I was attempting to use this particular
method of measurement (a) because background interference is automatically
accounted for in the way the PSD operates and (b) it was a completely
independent way of measuring the response. It appears only to work for
smaller format devices which I suppose have lower total capacitance
than the larger devices. It is intruiging though...
The 'recipe' we use for pre-flooding is the standard one
recommended by Mike Lesser.
- With the detector at ambient pressure in air, DRY air
ONLY is used now (19/9/95),
- a period of UV flood using a Cd lamp (strong 228nm line) placed
10cm from the detector.
We have tried times from 5 minutes to 30 minutes without any noticable
change in the response to the UV flood,
- measure the response straight away if required,
- Pump the cryostat to 10-4,
- Cool the chip to operating temperature and measure the response
in the way out-lined above. Now a few hours after UV flooding.
Loral Data #1 - 24/05/95
Data from the current (June 1995) room temperature and cold (-140C)
measurements are on the next Loral page-
View Loral Page - 1
Loral Data #2 - 19/09/95
Data from the current (August 1995) room temperature and
cold (-140C) measurements are on the second Loral page-
View Loral Page - 2
Loral Data #3 - 28/03/96
Data from the current (March 1996) room temperature and
cold (-120C) measurements are on the third Loral page-
View Loral Page - 3
This is for the Passivated Platinum Film (PPtF) Device
Loral Data #4 - 23/1/97
Data from an Engineering chip to proof the RGO monochromator
rig.
View Loral Page - 4
This is for a Passivated Platinum Film (PPtF) Device
Life, dont talk to me about life, brain as big as a planet and they ask
you to open the door.... Life....
Don't you wish you were here....
I am... apo