Applies an offset to the Y channel of the CS100 servo--loop
Format: CS100_YOFF <offset>
Units: none
Range: --2048 < <offset> < 2047
Defaults: Defaults to 0 on power--up
Examples : CS100_YOFF 100
Comments: Issuing CS100_YOFF will only result in a change of state of the CS100 once external control has been enabled (command CS100_EXT). The etalon will only change state if the CS100 has been switched to the etalon (command ETALON_VOLTAGE) and the servo loop closed (command CS100_CLOSE).