Preliminaries
- Switch off lights on all floors, close the blind in the restroom.
- In the control room, first check the Operations Logbook for any
restrictions that may be still in force. Turn on the DOME LIGHTS from the
console and raise the control room blind.
- Make a visual inspection of the observing floor (and the balcony above the
control room if the telescope is at ACCESS PARK) to check that the telescope
is free to move e.g. no ladders or steps are close to the telescope or
test instruments connected.
- Put the Telescope in ENGINEERING MODE by pressing the ENG/COMP push button.
The button should now be illuminated with an yellow light, indicating ENG mode
has been selected.
- Clear any emergency overrides by pressing the RESET button close to the
ENG/COMP button.
- Ensure that none of the SELECT MANUAL CONTROLS buttons have been
activated on the console
- Make sure that the TCS/DAS/ICL software is running. If not, please
follow this link before proceeding.
- Turn on OIL PUMP from the engineering rack.
Filling the WFC cryostat
- Moving the telescope to ACCESS PARK
To do this, on the engineering rack:
- Ensure the telescope is at ZENITH PARK and the HA synchro
pointers are at the "12:00" position.
- Turn the DEC QM variable speed control fully clockwise.
- Push the DEC QM button to release the slow motion clamps.
- Push and hold the DEC [+] button. This will move the telescope down
towards the balcony.
- At declination 100 degrees the HORIZON LIMIT alarm activates. The
telescope will stop and the slow motion clamps will re-engage.
- Push the ACCEPT button to mute the alarm.
- In order to move to 106 degrees (position for filling the cryostat)
turn and hold the LIMIT keyswitch clockwise in the OR
(override) position whilst pushing the DEC QM button
to release the slow motion clamps.
- Push the DEC [+] button until close to 105 degrees, the telescope will
then stop at about 106 degrees.
- Filling Cryostat
- When the telescope is at ACCESS PARK, wheel the N2 dewar close to the
platform.
- Connect the dewar to earth; use the earthing point on the hand rail to the
right of the telescope.
- FULLY insert the filler tube into the cryostat.
- On the dewar, CLOSE the PRESSURE RELEASE valve.
- OPEN the DECANT valve.
- OPEN the PRESSURE AUTO valve.
n.b. Check the pressure does not exceed 15 psi.
- If the dewar is low on liquid nitrogen, the PRESSURE MANUAL valve can be
opened. Use precaution when doing this as the pressure can build
up very quickly if the dewar is close to full.
- Wait until liquid nitrogen is SEEN to be coming out of the
cryostat.
- CLOSE the DECANT valve and the PRESSURE valve(s).
- OPEN the PRESSURE RELEASE. Removing the filler tube with care.
VERY IMPORTANT!
Wear safety helmet (with visor) and gloves when filling cryostats!
Always remember to leave the PRESSURE RELEASE valve on the dewar
OPEN. If this is forgotten, pressure can build up over time and burst
the relief valve on the dewar. This is especially true if the dewar is close
to full.
- Moving the telescope out of ACCESS PARK
To do this, on the engineering rack:
- Turn the DEC QM variable speed control fully clockwise.
- Turn and hold the LIMIT keyswitch clockwise in the OR
(override) position whilst pushing the DEC QM button to
release the slow motion clamps.
- When the clamps have disengaged, still holding the LIMIT keyswitch
to OR, push the DEC [-] button. This will move the telescope up and
away from the balcony.
- When declination is below 90 degrees you may release the keyswitch.
If the audible alarm sounds, push the ACCEPT button to mute it.
If the encoders have already been zeroset, push the DEC SM button and
put the TCS into Computer Mode. Bring telescope to zenith by USER>park zen.
If no zeroset has been done yet then,
- Continue to push the DEC [-] button until the telescope is vertical. The
DEC synchro pointers should be close to the "9:45" position.
- When the telescope is vertical, push the DEC SM button to re-engage the
slow motion clamps.
Zeroset encoders
- Visually check that the telescope is near zenith park (i.e.
vertical). The HA synchro position indicator pointers should be near the
"12 O`Clock" position with the DEC indicator pointers near the "9:45"
position. If the position indicators are some way off, move the telescope
close to these positions using the HA and DEC [+]/[-] buttons on the
engineering rack. Select quick motion (QM) if the telescope is a long way off
position then go to slow motion (SM) to make the final adjustments.
- In the TCS User window, enter:
USER>zeroset ha target (or: zero ha tar)
USER>zeroset dec target (or: zero dec tar)
- Using the HA and DEC [+]/[-] buttons with slow motion (SM) selected and
the drive rate controls adjusted for a very slow speed, move the telescope so
that the synchro dial pointers MOVE THROUGH the "12 O' Clock" position in HA
and the "9:45" position in DEC.
The TCS will beep if the zeroset is successful. The yellow LED's mounted
above the synchro dials will momentarily flash also confirming that the
telescope has been zeroset.
Opening up
- Opening the dome top shutter
On the engineering rack:
- Use RAISE MICRO until the OVERTRAVEL indicator goes off.
- Push RAISE MAIN until the shutter stops moving. The shutter is now fully
open.
n.b. If the dome or shutter does NOT work from the engineering rack
push buttons, go up to the balcony and check the manual controls on the large
grey cabinet mounted on the dome. The switches (key operated) associated with
the dome and shutters must all be pointing to REMOTE, not to LOCAL.
To operate the lower shutter, follow these
instructions.
- Push and hold the PRIMARY MIRROR OPEN button. Check the mirror petals open
fully (petals can be seen moving through a gap in the telescope structure).
- Push the DOME VENT FAN button to enable dome ventilation. The red light
will go OFF when the ventilator is on.
- Turn off the DOME LIGHTS from the console and close the control room blind.
Switching TCS in computer mode
- Put the telescope into Computer Mode by turning the COMP keyswitch. The
green section of the ENG/COMP push button switch should illuminate showing
that Computer Mode has been selected.
n.b. The HA and DEC slow motion (SM) push buttons MUST be
set for SM mode (i.e. All the lower green sections of the buttons
illuminated) before Computer Mode can be selected. If the yellow section of a
QM button is illuminated, push the adjacent SM button to re-engage the slow
motion clamps. Computer Mode should now be available.
TELESCOPE IS READY FOR OBSERVATIONS
Closing down
- Bring the Telescope to zenith by typing USER> park zen
- Push and hold the PRIMARY MIRROR CLOSE button. Check the mirror petals
close fully (petals can be seen moving through a gap in the telescope
structure).
- Push the DOME VENT FAN STOP button to turn off dome ventilation. The red
light will go on when ventilator is stopped.
- Closing the dome top shutter
Always remember to bring down the lower
shutter first
On SELECT MANUAL CONTROLS activate the button labeled dome/shutter.
On the engineering rack:
- Push LOWER MICRO until the MAIN OVER TRAVEL light goes off.
- Use LOWER MAIN until the shutter stops moving.
- Push LOWER MICRO to fully close the shutter.
Release the button labeled dome/shutter on SELECT MANUAL CONTROLS.
n.b. There are a few rules concerning dome shutter and bad weather.
When windy and raining from behind the shutter water can penetrate into
the dome; there is a seal but it isn't perfect. Therefore in this case
please turn the dome to point the shutter into the wind. On the other hand
with temperatures below zero we would like to avoid the build up of ice on
the shutter, in this case point the shutter out of the wind.
Switching TCS in engineering mode
- Put the Telescope in ENGINEERING MODE by pressing the ENG/COMP push
button. The button should now be illuminated with an yellow light,
indicating ENG mode has been selected.
- If an audial alarm sounds, type
USER>ack
to mute the alarm.
Filling cryostat - morning
As at the start of the night, the WFC cryostat needs to be filled:
- Bring the Telescope to Access Park.
- Fill the Cryostat. These procedures are described here.
- Turn off OIL PUMP from the engineering rack.
Writing tapes
There is a possibility that torn images are related to
the writing of tapes while doing exposures (and the consequent reading of
the chips). For this reason the following advice:
Please do not write tapes while doing observations; start writing them
in the morning when observations are finished.
- Check whether the drives need cleaning, this is the case if a LED is
flashing every 4 seconds. The cleaning cartridge is placed together with
the empty tapes together with instructions on how to use it.
- Insert the DDS3 (125m) tape in the drive /dev/rmt/0n or /dev/rmt/1n
- Logon to lpss8 as intguest and change directory to the data directory
/obsdata/...
- lpss8>fitsinit - follow the displayed instructions on how to
initialize the tape.
- lpss8>fitsout - again follow the instructions on the screen.
On a DDS3 tape (12GB) should be enough space for 150 runs. Note that a
list of missing runs will appear on the screen which you have to
acknowledge before the writing process begins.
Printing observing log
Fill in the observing log html form, select how many printed copies you need
and submit. Please archive one copy of the observing log in our log folder.
URL:
http://scorpio.roque.ing.iac.es:9999/intlogentry.html
Filling logbook
The printed logbook in the control room contains fields which must be
filled out during the night:
- CCD cryostats : Initials and local time for filling the CCD(s).
- Breakdown of observing time: Fill in any downtime that was encountered
(bad weather, technical problems, etc.) and note the time lost.
|