Home · Search · About ING · Astronomy · Public Information · Engineering

Security Information

  1. Please use the lone worker alarm when you are observing alone at the INT
  2. For security reasons you must follow the instructions given to you by the support astronomer, duty engineer or WHT telescope operator (eg. abandoning site in bad weather or regarding the operation of the telescope and instrument)

Preliminaries

  1. Switch off lights on all floors, close the blind in the restroom.

  2. In the control room, first check the Operations Logbook for any restrictions that may be still in force. Turn on the DOME LIGHTS from the console and raise the control room blind.

  3. Make a visual inspection of the observing floor (and the balcony above the control room if the telescope is at ACCESS PARK) to check that the telescope is free to move e.g. no ladders or steps are close to the telescope or test instruments connected.

  4. Clear any emergency overrides by pressing the RESET button close to the ENG/COMP button.

  5. Turn on OIL PUMP from the engineering rack.

Filling the WFC cryostat

Moving the telescope to ACCESS PARK (Engineering Rack)

  1. Ensure the telescope HA synchro pointers are at the "12:00" position (HA=0h).
  2. Turn the DEC QM variable speed control fully clockwise.
  3. Push the DEC QM button to release the slow motion clamps (engage Quick Motion).
  4. Push and hold the DEC [+] button. This will move the telescope down towards the balcony.
  5. At declination 100 degrees the HORIZON LIMIT alarm activates and the telescope will stop.
  6. Push the ACCEPT button to mute the alarm.
  7. In order to move to 106 degrees (position for filling the cryostat) turn and hold the LIMIT key-switch clockwise in the OR (override) position following the next few steps. The LIMIT key-switch causes the slow motion clamps to re-engage.
  8. Push the DEC QM button to go back to Quick Motion.
  9. Turn the DEC QM variable speed control to about half-way.
  10. Push and hold the DEC [+] button; the telescope will then stop at about 106 degrees by itself; there is no danger in keeping the DEC [+] button pressed as the QM rate is half-way.
  11. Now you may release the LIMIT key-switch.

Filling Cryostat (Balcony)

  1. IMPORTANT! Wear safety helmet (with visor) and gloves when filling cryostats!
  2. When the telescope is at ACCESS PARK, wheel the N2 dewar close to the platform.
  3. Connect the dewar to earth; use the earthing point on the hand rail to the right of the telescope.
  4. FULLY insert the filler tube into the cryostat.
  5. On the dewar, CLOSE the VENT valve.
  6. OPEN the DECANT valve.
  7. OPEN the PRESSURE AUTO valve.
    n.b. Check the pressure does not exceed 20 psi.
  8. If the dewar is low on liquid nitrogen, the PRESSURE MANUAL valve can be opened. Use precaution when doing this as the pressure can build up very quickly if the dewar is close to full.
  9. Wait until liquid nitrogen is SEEN to be coming out of the cryostat.
  10. CLOSE the DECANT valve and the PRESSURE valve(s).
  11. IMPORTANT! OPEN the VENT valve. If this is forgotten, pressure can build up over time and burst the relief valve on the dewar.
  12. Wait for the filler tube to warm up (until it is not rigid anymore) and remove it from the cryostat.
  13. Note down when crostat was filled in the Logbook.

Moving the telescope out of ACCESS PARK (Engineering Rack)

  1. Turn the DEC QM variable speed control fully clockwise.
  2. Turn and hold the LIMIT key-switch clockwise in the OR (override) position following the next few steps. The LIMIT key-switch causes the slow motion clamps to re-engage.
  3. Push the DEC QM button to go back to Quick Motion.
  4. Push the DEC [-] button. This will move the telescope up and away from the balcony.
  5. When declination is below 90 degrees you may release the LIMIT key-switch. If the audible alarm sounds, push the ACCEPT button to mute it.
  6. Continue to push the DEC [-] button and stop at DEC=28d 45' (+/- 1 degree). This needs a bit of experimenting because the telescope will not stop immediately when you release the DEC [-] button. The telescope is now vertical.
  7. Push the DEC SM button to re-engage the slow motion clamps.

Zeroset Encoders

  1. Visually check that the telescope is near zenith park (i.e. vertical). The HA synchro position indicator pointers should be near the "12 O`Clock" position (HA=0h) and the DEC indicator pointers near the "9:50" position (DEC=29d). If the position indicators are some way off, move the telescope close to these positions using the HA and DEC [+]/[-] buttons on the engineering rack. Select quick motion (QM) if the telescope is a long way off position then go to slow motion (SM) to make the final adjustments. Both QM and SM buttons will be lit when SM is activated.

  2. In the TCS User window, enter:
    USER> zeroset ha target  (or: zero ha tar)
    USER> zeroset dec target (or: zero dec tar)
        
  3. n.b. These commands need to be typed in pair. If you type only one and then zeroset only that axis, it won't work. Using the HA and DEC [+]/[-] buttons with slow motion (SM) selected and the SM variable speed control to about half-way, move the telescope so that the synchro dial pointers MOVE THROUGH the zeroset marks in both axes.

    The TCS will beep if the zeroset is successful. The yellow LED's mounted above the synchro dials will momentarily flash also confirming that the telescope has been zeroset.

    Note: Occasionally when moving through the zeroset position the yellow LED will flash but the TCS computer does not beep to acknowledge the zeroset. If this happens, move back through the zeroset position and try again selecting a slower speed with the drive rate controls. If this also does not help then re-issue the zeroset commands on the TCS for both axes and repeat moving the telescope through the zeroset positions.

Opening Up

  1. Check that the weather conditions allows for observations. Set the humidity audible alarm to either 75% or 90%.

  2. On the UPPER SHUTTER PANEL, use RAISE MICRO until the MAIN OVERTRAVEL light underneath turns off.

  3. If sunlight is seen hitting the inside of the dome, rotate the dome away from the sun using the DOME ROTATION cw (clockwise) and ccw (counter clockwise) buttons found just right of the upper shutter panel.

  4. On the UPPER SHUTTER PANEL, push RAISE MAIN until the shutter stops moving.

  5. Push and hold the PRIMARY MIRROR OPEN button. Check the mirror petals open fully (petals can be seen moving through the gap in the telescope structure).

  6. Push the DOME VENT FAN button to enable dome ventilation. The red light will go OFF when the ventilator is on. You can also open the 'south doors' provided the TO did show it to you. Keep the south doors closed if the wind speed is higher than 40 km/h. Should the weather turn bad it is your responsibility to close this doors immediately.

  7. Turn off the DOME LIGHTS from the console, the low-level dome lights (wall switch) and close the control room blind.

Switching TCS in computer mode

  1. Put the telescope into Computer Mode by turning the COMP key-switch. The green section of the ENG/COMP push button switch should illuminate showing that Computer Mode has been selected. The system only switches to computer mode if the OIL PUMP is on, no emergency override is active (press the RESET button close to the ENG/COMP button) and if HA and DEC slow motion (SM) are active.

Initial TCS commands

USER>refers to the TCS User Interface window
SYS>refers to the pink Instrument Control System window

USER> show focalShow the focal station setup. If not set to PRIME/WFC, run the command SYS>startobssys.
USER> focus 43.98Set focus to a 'good' value.
USER> calibrate lastLoad the last TCS pointing model.
USER> out termList the objects loaded into the TCS.
USER> incl userAdd objects by loading a predefined catalog. This assumes that a catalog named 'user.cat' is stored in /int/cat.
USER> incl newblank38add WFC blank fields. See our table of blank fields.
SYS> xhost +Open a copy of the TCS info display on the observing system screen.
USER> clone on 161.72.6.9

Setting up the Autoguider

  1. SYS>cmd INFODISP PREDICT ON
    This enables the automatic procedure for finding guide stars. Only valid for a rotator position of 180 degrees and nominal pointing (center of CCD 4). Every gocat command will now list a set of suitable guide stars in the TALKER window.

  2. From the autoguider menu, select an integration time of 1 second and a guide window of 150 pixels (~40 arcseconds).

Twilight Sky-flats

There is an automatic procedure which, by continually reading the auto-guider chip, tells the observer what exposure time to use in order to achieve about 20000 counts. Suitable blank fields are listed above the TCS terminal, together with instructions on how to include them in the catalog.

  1. SYS> gocat xxxxxx (chose a suitable blank field)

  2. SYS> guide on pixel 1032 512
    The guider is only used to measure the sky brightness, we don't want to guide. To open the guide loop click the 'open loop' button on the AG GUI. If the telescope is GUIDING and this button shows 'close loop', click on it and then click again to open the loop.

  3. SYS> smartflat &
    A dialog box will pop up and tell the observer the needed exposure time. Exposures are taken by either clicking the `take this flat' button or using the command line, eg. SYS>sky 10 "Skyflat I 10s"
    Smartflat needs the camera shutter open!

  4. Apply telescope offsets between successive exposures to enable removal of objects on the masterflat by i.e. median filtering: SYS>offset arc 10 10 (absolute offsets)

  5. Once all twilight flats are done:

    click on the 'STOP' button on the autoguider GUI.

    Accept any error box messages and close the smartflat window.

Single Star Calibrate

This is the easiest procedure to check and readjust the telescope pointing. If the star do not appear at all, redo the zeroset procedure. If the pointing seems bad even after going through this procedure, please refer to the troubleshooting section on how to perform a 7 Star calibrate.

SYS> agwin acq pos 1032 512Readout the auto-guider centre.
USER> cal faintThis sends the telescope to a suitable calibrate star.
When the HANDSET menu appearshit <ctrl-Z> to break out.
USER> aper 5Define telescope pointing on the auto-guider centre.
On the auto-guidermark the window centre by using the menu item SETUP/TOGGLE CENTRE CROSS.
SYS> autotv onTurn on continuous readout mode.

If the star appears close to the window centre, the pointing is ok and you can skip the rest of this section. If not use the HANDSET to centre the star on the centre cross (in XY mode for the 180 deg. rotator position the cursors are reversed). Once centred type:

USER> point calibrate
USER> cal anal zero

The rms of this solution is typically 0.01", type "Y" to accept.

Focusing the Telescope

To determine the best telescope focus select a suitable field (eg. one of the WFC photometric standard fields) and run the focus script:

  1. SYS> gocat xxxxxx (chose a suitable standard)

  2. SYS>focusrun &

    A dialog box will pop up and the appropriate values may be filled in and a focusrun started. All "#n" exposures will be taken on the same frame with the a double spacing between #1 and #2.

    Remember to set focus to the best value at end of focusrun (e.g. SYS>focus 43.98) and to repeat gocat if you need the target back in the centre.

  3. Accept any error box messages and close the focusrun window.

  4. You can use the Iraf command "wfc_starfocus" to compute the best focus value from the focusrun.

Observing

Here it follows a list of the most commonly used commands throughout the night.
Remember that the lower shutter starts vignetting at ZENITH DISTANCE >57 degrees (ELEVATION <33 degrees) - the system doesn't warn you! If you need to observe at ZENITH DISTANCE > 57 degrees, please follow the instructions described in the troubleshooting section.

Object Handling:

USER> incl 'user'Includes a user catalog named user.cat, stored in /int/cat
USER> out termList objects currently loaded into the TCS (but not the system catalog)
SYS> object name 12 20 30 -10 5 40 J2000
SYS> addAdd an object to the catalog
USER> output file int:[cat]virgoWrite virgo.cat back to disk

Acquisition and Guiding:

SYS> gocat N6720If loaded into the TCS, moves the telescope to an object called N6720
SYS> gocat 92-248Moves the telescope to the landolt field 92-248 stored in the system catalog
SYS> offset arc -5 5Offsets the telescope by -5 arcsec in RA and +5 arcsec in Dec from nominal pointing (absolute offset). If it fails, Ctrl-C and issue the same command at the USER> prompt or use:
SYS> cmd TCS USER "OFFSET ARC -5 5"The same but they say this doesn't fail!
USER> TWEAK -5 5 0XY Offset keeping the guideloop closed
SYS> cmd TCS USER "TWEAK -5 5 0"Same command invoked from the observers workstation
SYS> guide on pixel x yStarts guiding on pixel x, y. The values are chosen from one of the guide stars listed on the TALKER window. The telescope will be moved until the guide star appears in the center of the guide-box
SYS> agwin gui pos x yPositions the guide window within the autoguider active area, 'guide on' will now lock guiding without centering.
SYS> agwin acq pos x yPositions the acquisition field within the autoguider active area.
SYS> fieldDo a field and search for potential guide objects from current position
SYS> guide on star nPick "n"th brightest detected object and guide.
SYS> guide [on|off]Turn guiding on/off
SYS> autotv [on|off]Turn continuous readout mode on/off

Exposing:

SYS> run 1200 "N6720"Expose for 1200 seconds and name the image N6720
SYS> multrun 3 120 "N6720"Take 3 consecutive exposures of 120 seconds each
SYS> biasTake a bias frame
SYS> flat 10A 10 second exposure with FITS header keyword OBSTYPE set to 'flat'

Filters:

SYS> filter RChanges filter to R. Alternatively, the GUI can be used to change filter.

Examining data:

The IRAF interface on the Data Reduction System is used to examine the data. From the IRAF prompt, make sure you are in the right directory, i.e. cl>cd /obsdata/inte/20020319

cl> display r242645[4] 1Loads the content of CCD #4 into buffer 1 and displays it in Ximtool.
cl> wfc_display r242645Loads each of the 4 CCDs into 4 different buffers and displays them all in Ximtool. The 'tile frames' options in Ximtool must be activated.

To access pipeline reduced images, replace r123456.fit with qldata$q123456.fit

Closing down at end of night

  1. Bring the Telescope close to access park by typing:

    USER> park ap1

  2. Push and hold the PRIMARY MIRROR CLOSE button. Check the mirror petals close fully (petals can be seen moving through a gap in the telescope structure).

  3. Push the DOME VENT FAN STOP button to turn off dome ventilation. The red light will go on when ventilator is stopped. Close the 'south doors' when opened before.

  4. Once the telescope is STOPPED, switch the TCS to engineering mode by typing:

    USER> eng

    The ENG/COMP push button should now be illuminated with an yellow light.

  5. Close the dome shutter pushing LOWER MAIN until the shutter stops moving.

  6. Push LOWER MICRO to fully close the shutter.

  7. Fill the WFC cryostat following the same procedures as at the start of the night. Leave the telescope at access park.

  8. Turn off OIL PUMP from the engineering rack. The OIL alarm goes off after a while; just push the ACCEPT button.

Writing Tapes

  1. Insert the DDS3 (125m) tape in the drive /dev/rmt/0n or /dev/rmt/1n. A DDS3 tape (12GB) should be able to store 150 runs in fits format, or about 300 using tar.

  2. Open a new xterm on the Data Reduction machine and change directory to where the data are stored.

  3. prompt>tar cvf /dev/rmt/0n *.fit - writes all fit files to tape.

Printing Observing Log

Fill in the nightlog which is accessed from the observing logger window. Submit and print the copies you need. Please archive one copy of the observing log in the log folder in the control room.

Filling the Logbook

The printed logbook in the control room contains fields which must be filled out during the night:
  1. CCD cryostats : Initials and local time for filling the CCD(s).

  2. Breakdown of observing time: Fill in any down-time that was encountered (bad weather, technical problems, etc.) and note the time lost.

Leaving the Building

Before leaving the building, please switch off all mayor lights in the control room and in the kitchen area. Check that no electrical devices are left on in the kitchen, i.e. sandwich toaster, coffee machine, etc.

Observers hand over check

Ready to go it alone? Please check you are familiar with the following tasks:


ING Logo
Last Updated: 24 March 2004
Jürg Rey juerg@ing.iac.es