Previous: Target acquisition with a multiaperture mask
Up: Target acquisition with a multiaperture mask
Next: Further refinements to acquisition
Previous Page: Target acquisition with a multiaperture mask
Next Page: Further refinements to acquisition

Standard technique

The standard procedure is summarized in Fig.10.

Fig. 10

  1. Select the required position angle

    The first thing to do is to select the position angle that you require which should obviously be the same as you specified when designing the mask. The Telescope Operator will use the TCS command ROT/SKY to set this up. If the TCS has been setup correctly for LDSS-2, it will put the telescope rotator at the angle which puts the LDSS-2 slits at the position angle requested.

  2. Local calibration of the pointing model with a fiducial star

    The acquisition procedure requires the telescope to point to an accuracy that is a small fraction of the size of the reference star holes in the mask. Since there may be an offset between the coordinate system of the telescope (FK5) and that of the target positions, it is very wise to refine the pointing model by moving the telescope to a star near (or in) the target field which has astrometry from the same solution as your target objects - the pointing recalibration star defined in Section 2.2. This star should be loaded into the TCS catalogue and acquired using GOCAT, and then moved using the handset onto the acquisition reference position marked on the TV display.

  3. Blind offset to the target field

    The TCS command BLIND_OFFSET target will move the telescope to the position of target on the assumption that the position of the original star is perfectly known. In other words it locally refines the pointing mode. In all likelihood the telescope will move to target with sub-arcsecond accuracy, provided that the local reference star was very close to the target field. Usually this reference star will be one of the reference stars in the target field.

  4. Acquisition of the guide star

    Now that you are on the target, it is time to find a guide star. The simplest way is to use the GSS software on LPVS3 which will provide a list of suitable stars with the required guide probe positions. Once a star is found, don't start guiding yet but repeat the blind offset sequence (i.e. GOCAT to your positional reference star, centre up and do the BLIND_OFFSET again) to make sure that telescope has not wandered while you were finding the guide star. Start guiding immediately you have done the blind offset, and as soon as you are sure that the autoguider is tracking the star, ask the Telescope Operator to make the telescope use the autoguider output for tracking corrections.

  5. Direct exposure through mask

    Now take a direct image of the sky through the mask with no grism (a broad band filter is desirable but not essential). The image should show the reference stars in the appropriate holes in the masks, in general offset from the centres of the holes. If no stars can be seen: check the acquisition procedure so far - are you still autoguiding? --- is the mask correct? An image of the target field without the mask will verify the field identification if in doubt. Measurements of the positions of the reference stars can be compared with the location of the holes in the masks to check for errors.

  6. Determination of the telescope offset required

    The image can now be analyzed using LEXT procedure FIND_TWEAK where is the number of fiducial holes. This works out how much the telescope should be moved to place the fiducial stars bang in the middle of the fiducial holes. Read the comments as you go along for information about the things that happen. The main stages are:

  7. Application of the telescope offset

    Give the offsets to the Telescope Operator who should use the TCS command:

    to move the telescope while the autoguider is still locked on and the telescope continues to receive guiding signals. During the set up procedures the Telescope Operator should have used the TCS command INSTRUMENT LDSS to enter the correct orientation for the offsets.

    This is because the autoguider may lose lock if the telescope moves by amounts larger than this. The procedure is to break down the TWEAK into increments of this size and apply each one, waiting a few seconds to allow the autoguider to catch up. It would be wise to watch the autoguider display carefully to try and see if it loses lock. If it does, you could try (a) reducing the increment size and/or (b) selecting a brighter guide star. You should also take care that the guide star is not so near the edge of the autoguider field that it moves outside the area in which the autoguider can track it when the TWEAK is applied.

    Doing the TWEAK with the autoguider in closed-loop should ensure that no positional errors occurs during the offset. If you need to apply a large TWEAK that cannot be done in closed-loop (for example, because the guide star would move outside the autoguider detector area), you can use TWEAK with the autoguider off ( open-loop). However you would then need to take another direct exposure after restarting autoguiding to check the position of the targets before doing a smaller closed-loop TWEAK.

  8. Acquisition check

    In general, this procedure will be sufficient to acquire the target field to an accuracy of around 0.1 arcsec but it must be checked. After performing the TWEAK, take another direct image through the mask. Before analyzing it, it is usually safe to put the grism in and start a spectroscopic exposure while the second direct exposure is analyzed. If there is a residual TWEAK to be done, this can be applied during the spectroscopic exposure (in closed-loop). If there is a large error in position, the exposure can be aborted and the whole procedure repeated. If this happens, you should first try and work out what has gone wrong.

From time to time, during a long sequence of spectroscopic observations on the same field, you may wish to check the acquisition. To do this, simply take a direct image of the sky through the mask and use FIND_TWEAK to check that the stars are still centred in the holes. Any fine adjustments can be made during the spectroscopic exposures.



Previous: Target acquisition with a multiaperture mask
Up: Target acquisition with a multiaperture mask
Next: Further refinements to acquisition
Previous Page: Target acquisition with a multiaperture mask
Next Page: Further refinements to acquisition


Wed Mar 16 00:17:46 GMT 1994