Investigation of the emissivity in K of INGRID
(introduction and working plan)
Oct 2003; T. Gregory, A. Zurita
===============================================
Introduction
------------
-The aim of this plan is to investigate the origin of the
unexpectedly high emissivity in K of INGRID when used
with NAOMI.
-Tests performed up to now suggest:
a/ there is a significant source of K emission which
seems to be located between the INGRID entrance
window and the pupil stops (with no foreoptics
and INGRID looking at a cold source, we measure some
900 adu/s at the detector).
b/ the emissivity in K with NAOMI foreoptics is
approx. double than with the Cass foreoptics.
-It was decided in a AO/INGRID meeting (Aug 03):
-change entrance window of INGRID (to ensure the material is CaF2)
-get the NAOMI foreoptics coated
Although these actions could reduce the background in K, the source
of high emissivity is still unknown (specially because of a/)
Plan before and during the INGRID stand-down
--------------------------------------------
A/ Confirm previous measurements
With INGRID cold, repeat experiments with the cold soup ladle
to measure the amount of K emissivity reaching the INGRID detector
with: no foreoptics, Cass foreoptics, NAOMI foreoptics.
o Requirements:
-INGRID must be cold
-Full access to INGRID (Test focal station or aluminising area)
-Both fore-optics available (Cass and NAOMI)
-INGRID should be rotated by 180 deg (upside down) to
simulate both (Cass and NAOMI) orientations (trolley needed ??)
o Total time needed : 2 days
B/ Examination of the baffling in front of the cold stops
Examination of the baffling: shape, colours and shininess.
The pupil stop baffle currently inside INGRID seems to
have a wrong shape (i.e. it could be allowing unwanted
radiation to reach the detector)
o Requirements: INGRID warm and open
o Total time needed : half a day
C/ Manufacturing and installation of a new Pupil stop baffle
From conversation with P. Jolley, it must be possible to
fit a new pupil stop baffle inside INGRID (to replace the
one mentioned in B/).
**The pupil stop baffle could be manufactured before INGRID
is opened up.**
o Suggested plan (assuming that several cycles (2 at least)
will take place):
-manufacture a new pupil stop baffle (a simple one: not polished
nor coated), before INGRID is opened.
-fit it inside INGRID at first opportunity
(before the the first cool down)
-test it (repeat A/) when INGRID is cold
-if it works, make a new pupil stop baffle (polished and gold
coated outside, black inside) and install inside INGRID when open.