High Level Requirements

 

Project name

Acam

 

Release

Draft

Date: 22nd July 2007               

 

PRINCE2

 

 

Co Authors:

 

Chris Benn / Kevin Dee

Owner:

 

Don Carlos

Client:

 

ING

Document Number:

 

ACAM_HL_REQ

 

 

 

 


Document History

Document Location

This document is only valid on the day it was printed.

Printed on July 2007.

The source of the document will be found at:

http://www.ing.iac.es/~eng/private/projects/ACAM/ACamweb/documents/current/High_Level_Requirements.htm

Revision History

Date of next revision:    ?

 

Revision date

Previous revision date

Summary of Changes

Changes marked

 

22 Jul 07

N/A

Document created & updated.

Kevin Dee

 

25 Jul 07

22 Jul 07

Requirements amended & updated

Chris Benn

Approvals

This document requires the following approvals.

Signed approval forms are filed in the project files.

 

Name

Signature

Title

Date of Issue

Version

 

Kevin Dee

 

Acam Project Manager

 

 

 

Chris Benn

 

Acam Project Scientist

 

 

 

Don Carlos

 

Head of Engineering ING

 

 

Distribution

This document has been distributed to:

 

 

Name

Title

Date of Issue

Version

 

 Kevin Dee

 ACam Project Manager

 22/07/2007

 Draft

 

 Chris Benn

 ACam Project Scientist

 22/07/2007

 Draft

 

 Don Carlos

 ING Head of Engineering

 22/07/2007

 Draft

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 


Purpose

 

 

To summarise all of the ACAM High level Requirements or Constraints and their status.

 

 

 

Reference

Last up dated

Requirement /Constraint

Optimum Value

Acceptable Value

Comments

Status

001

22-July-2007

Observing Modes

High Resolution/Throughput Imaging 350nm-1000nm.

Long Slit Spectroscopy. (1arcminute good PSF- make available up to 8 arcmin)

 

High Resolution/ Throughput Imaging 360nm-850nm

Wide field imaging is the primary mode in which the instrument must be optimised for.

Image quality 850-1000 nm is not critical but will probably be OK anyway.

 

 

Frozen.

High Resolution/Throughput Imaging 350nm-1000nm.

Low resolution Slit Spectroscopy

002

22-July-2007

Unvignetted Field of View (FOV) Diameter

9.3 Arcmin (ref 122.7 mm dia @ telescope focus)

8.3 Arcmin (ref 110.4 mm dia @ telescope focus)

A larger field up to 11.8 arcmin would make use of CCD corners and width on a 2k by 4k chip  but constraints in particular costs of optics and complexity of design and of equal importance  the weight of the optics – which in turn would effect the total weight constraint of the instrument force this value lower.

Frozen 8.3 arcmin

003

22-July-2007

Image Quality

Imaging Mode 80% encircled energy diameter

380 - 1000nm

<0.25” FOV r<4.15’

<0.5” FOV r>4.15

350 – 380nm

<0.25” FOV r <1.0’

<0.5” FOV 1.0< r <4.15’

<1” FOV r >4.15’

380 - 1000nm

<0.3” FOV r <4.15’

<1”  FOV r >4.15’

350 – 380nm

<0.35” FOV r <1.0’

<0.7” FOV 1.0< r <4.15’

<2” FOV r > 4.15’

 

 

 

The image quality is relaxed as a function of field size in the UV wavelengths, as the design complexity, cost for exotic glass types, number of lenses all increase to facilitate the UV. Overall throughput across the visible wavelengths should not be compromised to accommodate an improvement in image quality across a large FOV for UV.

 

?

004

22-July-2007

Spectral Resolution @ 600nm 0.75” slit

380 – 1000nm

R>1000 1 arcmin

R=>500 Full Field

350 – 380nm

R>500 1 arcmin

 

380 – 1000nm

R>500 1 arcmin

R>200 Full Field

350 – 380nm

R>250 1 arcmin

 

The key requirement is that the on-axis spectroscopic resolution in the visible range is attained. The resolution in the UV is less critical. In the UV, the spectroscopic resolution is poorer (because the aberrations are larger), but the impact of this on the science is small. Off-axis, the spectroscopic resolution is also degraded, but the impact of this on the science is again small, because most spectroscopy would be of single on-axis targets.

?

005

22-July-2007

Image Quality

Slit Spectroscopy Mode 80% encircled energy diameter

380 - 1000nm

< 2 pixels on axis

< 3 pixels full field

350 – 380nm

< 2 pixels on axis

< 4 pixels full field

380 - 1000nm

< 3 pixels on axis

< 4 pixels full field

350 – 380nm

< 3 pixels on axis

Full field not required

 

?

 006

22-July-2007

Slit

Slit slide

Slit length = 8 arcmin

Slit width arcsec  = 0.5, 0.75, 1, 1.25, 1.5, 1.75, 2, 10 & out

 

Slit slide

Slit length = 5 arcmin

Slit width arcsec  = 0.5, 0.75, 1, 2, 10 & out

 

At a minimum, one slit of size 1" x 8' should be provided, but ideally, a choice of slit widths should be available e.g. 0.5, 0.75, 1, 2, 10". These could be mounted on a slide, leaving open the future possibility of installing multi-slit masks.

Frozen.

3 Slit Positions available, mounted in instrument at any one time covering the 8.3 arc min field. 

Interchangeable Slit masks available = 0.5, 0.75, 1, 1.25, 1.5, 1.75, 2, 10  arc sec.

Commissioning and diagnostic multi holed mask to cover full 8.3 arcmin field.

 

 

 007

22-July-2007

Instrument Entrance aperture

9.3 arc min

8.3 arc min

Instrument entrance aperture to be fitted.

Frozen @ 8.3 arc min

 008

22-July-2007

Pupil Aperture

50mm

>40 mm <70mm

Larger pupils are worse for nb filters & increase the optical path length.

Small pupils increase the complexity of the design

 

 

Currently 45 mm diameter. Driven by optical design

 009

22-July-2007

Camera Focus

Re-focus of telescope secondary for filter changes and different wave lengths.

A focus translation stage in the actual camera is an acceptable option.

If there is a clear gain in image quality in both imaging and spectroscopic modes by using a focus stage in the camera – then this can be incorporated in the design.

 

Frozen.

Telescope re-focus required on filter changes.

Slit refocus TBC after spectroscopic design is frozen.

.010

22-July-2007

Camera Optical length I.e. from telescope focus (which is close to the outer diameter of the A&G box) to CCD.

>600 mm

<850 mm

Making the instrument length short reduces the effects on mechanical, thermal and optical stability. Distance between certain optics must be maintained to accommodate and optimise the use of a slit mask, filters, Grisms and a shutter.

TBC. Driven by Optical design

 011

22-July-2007

Pupil exit distance to next optical lens.

>50 mm

Half thickness of GRISM/VPH plus 7.5 mm.

NB filters & Grisms will sit at pupil. Optical design optimizes by brining next optical component close to pupil. Filter wheels & structure constrain this distance

TBA. Between Optical designer & Mechanical Engineer

 012

22-July-2007

Camera throughput

350-380 nm >0.8

380-1000nm >0.8

350-380 > 0.6

380-100nm >0.75

Barr indicates that to get an anti reflecting coating that covers 350-1000nm you may suffer losses up to < 2% a surface. If you only consider 400nm -1000nm you can reach <1%  a surface. CaF2 due to its low index basically act's as an AR coating all on its own and Barr do not recommend coating it.

?

 013

22-July-2007

Narrow band filters

Band Pass =>15 A

Allow for filter tilt to remove ghosting.

Standard.8=Number of filters available. Up to 80 mm dia. 10mm thick.

Special<8=Number of filters

up to 120 mm dia x 12 mm thick.

Band Pass =>15 A

Allow for filter tilt to remove ghosting.

Standard.6=Number of filters available. Up to 80 mm dia. 10mm thick.

Special.<6=Number of filters up to 120 mm dia x 12 mm thick.

Location: chosen to minimise the *range* of angles of incidence of incoming rays, to minimise the range of wavelength shifts. A constant wavelength shift across the whole field is OK, a difference between on-axis and edge of field is not.  Astron's design the wavelength shift between on-axis and edge of field happens to be ~ 5 A (i.e. ~ acceptable) for nband filters both before *and* after L2/L3, but this may change with further optimisation

Frozen.

Location @ camera pupil.

Standard diameter of filters. 76 mm.

Standard number of filters positions in instrument. 5 + 1 clear + 1 VPH.

6 position combinations of 76 mm NB filters & VPH/GRISMs + 1 clear

Specials up to 110 Diameter.

 014

22-July-2007

Broad Band Filters

Position in front of NB filters.

8 = Number of filters

U B V R I Z clear

80 mm diameter 5 mm thick

Position in front of NB filters.

7 = Number of filters

U B V R I Z clear

80 mm diameter 5 mm thick

Location: not critical, but ideally before the narrow-band filters, to remove bulk of energy from beam before it gets scattered by the many layers in the narrow-band filter. Positions: 7, almost always with U B V R I Z filters. Diameter of filters: slightly larger than beam, to allow for tilting to remove ghosts. The current 50-mm ones are almost large enough for Astron's design.

Frozen.

Up stream of Narrow band filters.

Filter Diameters 76mm.

6 filters + 1 clear.

 015

22-July-2007

VPH/ Grisms in Littrow Configuration.

Same position as NB filters.

8 = number of Grisms

76 mm diameter

thickness > 20 < 30

 

Same position as NB filters.

6 = number of Grisms

Diameter plus 5mm on pupil diameter

thickness < 50

 

VPH/Grisms which exploit the optimized imaging instrument

TBC

 016

22-July-2007

Order sorting Filters

Filter slide.

Position in front of Broad band filters.

Number of filters 5

Filter slide.

Position in front of Broad band filters.

Number of filters 5

 

Frozen.

To be included with VPH/ grism  for a standard set up

 017

22-July-2007

CAGB ACam science fold mirror

Size = 11 arcmin .

Elliptical mirror size

Minor axis = 190  Major axis = 270

coating optimised for λ range

Surface finish  λ/20

Size = 10 arcmin .

Elliptical mirror size

Minor axis = 190  Major axis = 270

coating Aluminium

Surface finish  λ/20

 

 

New coatings to be investigated. Mirror size set at 11 arcmins which is constrained by the autoguider Patrol field

Frozen @ 11 arcmins

 018

22-July-2007

Patrol Field

Keep current patrol field

Maximise fold flat with out compromising ability to find adequate guide stars

Ref. Document : CAGB autoguider patrol field and fold flat.pdf

Frozen @ 20% lose of original Patrol field.

 019

22-July-2007

Calibration source

Use current CAGB unit.

Fold flat to feed light on axis into ACam is mounted on an off axis rotation stage combined with the CAGB science fold mirror

Use current CAGB unit.

Fold flat to feed light on axis into ACam is mounted on an off axis rotation stage combined with the CAGB science fold mirror

This has been modelled and provides a cost effective and efficient calibration source for ACam spectroscopy

Frozen.

Utilise current small auxport fold flat

 020

22-July-2007

Slit Acquisition

It may be feasible using fold mirrors and some optics to use the current acquisition TV camera pick up the back reflected light of the slit unit which is on a 10 degree angle.

The easiest & cheapest option for target acquisition is to view the field in imaging mode, position the target on a predetermined pixel, then switch to spectroscopic mode.

 

Fixed.

Target acquisition is to view the field in imaging mode, position the target on a predetermined pixel, then switch to spectroscopic mode.

 021

22-July-2007

Camera Plate Scale

0.25”/pixel

0.25”/pixel

Provides good sampling for best seeing 0.5”

Fixed @ 0.25”/pixel

 022

22-July-2007

CCD

E2V 2048x4100x15micron pixels.

 

Marconi.

2047 x 4611 x 13.5 micron pixels.

E2V same QE curve and fringe performance as the ISIS Red+ detector. The central 2kx2k pixels can be windowed off to improve readout speed.

Fixed.  E2V order complete. Delivery July 2007.

 023

22-July-2007

CCD Fringing

+/- 1% at 800nm

 

 

 

 

+/- 1% at 800nm

E2V meets fringing requirements

 ?

 024

22-July-2007

CCD cryostat window

Flat window

Distance from window to CCD >10mm< 12mm.

Distance from window to preceding optical element. >10 mm

Material should be fused silica to avoid background radiation from the glass

Active Window.

Distance from window to CCD >10mm< 12mm.

Distance from window to preceding optical element. >7.5 mm .

Material should be fused silica to avoid background radiation from the glass

It is preferred to have a flat cryostat window in the optical design to enable the CCD to be used at any focal station. This also means a standard mounting arrangement can also be used. Image quality must meet requirements; if this is not the case then an active window should be incorporated in the optical design. Note consideration for the effects of vacuum and atmospheric pressure across the window need to be addressed.

Frozen. Active cryostat window.

Distance from optical edge to CCD

= >7.5mm

 025

22-July-2007

Weight of instrument

Out of Balance Moment

80 kg

< 100Nm

110 kg

< 150 Nm

Instrument must be counter balanced to reduce the out of balance torque on the rotating cass cluster.

TBC

 

 026

22-July-2007

Instrument Interface

. Open aperture flange is 220mmÆ. Fixing 12 x M6 equispaced on a PCD of 230mmÆ.

Current auxport interface flange. Open aperture flange is 220mmÆ. Fixing 12 x M6 equispaced on a PCD of 230mmÆ.

Current mount can be stiffened to minimise flexure.

Fixed. Current CAGB auxport interface flange

 027

22-July-2007

Instrument Space Envelope

Optical axis and interface flange as datum. 600 mm (long) x 450 mm wide x 450 mm (high)

Optical axis and interface flange as datum. 850mm (long) x 600mm wide x 600mm (high)

This excludes the actual cryostat. NB. Electronics etc. can be mounted outside the space envelope after considerations for collisions and outer balance forces have been taken into account.

 TBC

028 22-July-2007 Focal plane Deflection < 25 microns from optical axis at CCD in any observing orientation. < 30 microns from optical axis at CCD in any observing orientation. First pass flexure analysis indicates instrument structure can deliver 30 microns deflection through a 90 degree gravity vector change. More detailed analysis required, including, thermal, vibration and shock. TBC.
029 22-July-2007 Focal Plane Tilt < 0.06 mill-radians tilt from focal plane at any observing orientation < 0.1 milli-radians tilt from focal plane at any observing orientation First pass flexure analysis indicates instrument structure can deliver 12 arcsec tilt through a 90 degree gravity vector change. More detailed analysis required. TBC
030 22-July-2007 Shutter 100m diameter clear aperture prontor shutter - modified for air operation. Position after Pupil and next lens element. 100m diameter clear aperture prontor shutter - modified for air operation. Position up stream of Broad Band filters. Relative large aperture allows the shutter to be fitted in many positions in the camera. Instrument is designed to be light tight - essential down stream of shutter. Frozen. Air operated. Up stream of Broad band filters